Operation – KROHNE OPTISENS COND 1200 EN User Manual

Page 26

background image

4

OPERATION

26

OPTISENS COND 1200

www.krohne.com

01/2014 - 4001967301 - MA OPTISENS COND 1200 R01 en

C1, process input A

C1, process input A

C1, process input A

C1, process input A
C2, process input B

C2, process input B

C2, process input B

C2, process input B

Level

Designation / function

Settings / descriptions

Process input A and B can be either a sensor 1 or a sensor 2. Further information about the type of sensor 1 or 2 please

refer to MAC 100 manual "Sensor input combinations". Process input A is always present, process Input B can be

present.

Note: The exchange of a sensor 1 with a sensor 2, or vice versa, can only be done by the manufacturer!

Depending on the sensor which is connected to a slot A or B the menu changes.

C1.1

C1.1

C1.1

C1.1

parameter (conductivity)

This menu item is for selecting the probe which is connected to process

input A/B. The entries of this selection depends on the chosen device

configuration. The device configuration is customer specific and set during

production.

C1.2

C1.2

C1.2

C1.2

cell constant

Enter cell constant.

C1.3

C1.3

C1.3

C1.3

cable resistance

Enter cable resistance in Ohm

C1.14

C1.14

C1.14

C1.14

time constant

Enter time constant.

C1.15

C1.15

C1.15

C1.15

temperature

Menu for temperature measurement. Available for sensor 1 and sensor 2.

C1.15.1

probe

Options:

manual: used if no internal or external temperature sensor is connected to
the signal converter

Pt1000: used if an external Pt1000 temperature sensor is connected to the
signal converter

C1.15.2

manual

Only available if C1.15.1 is set to "manual". If you have chosen "manual"

enter temperature.

C1.15.3

correction

Offset correction for temperature measurement. Not available if C1.15.1 is

set to "manual". If you have chosen “Pt 1000” enter the temperature

correction.

C1.15.4

limitation

Measuring ranges for temperature measurement. Enter limitation

C1.15.5

temp. comp.

Menu for editing the temperature compensation parameters for the

measurement.

Options:

off

table

ultra pure water NACl

ultra pure water Nao3

upw morpholin

ultra pure water NH3

ultra pure water HCL

ultra pure water

linear

C1.15.6

temp. coefficient

If you have chosen “linear” enter the temperature coefficient.

C1.15.7

ref. temperature

If you have chosen “linear” enter the reference temperature.

C1.15.8

Reference points

Only available if C1.15.5 is set to “table”. Choose:

2 till 10

C1.15.9

tbl. temperature

Only available if C1.15.5 is set to “table”. Choose:

temperature for each row

C1.15.10

tbl. conductivity

Only available if C1.15.5 is set to “table”. Choose:

conductivity for each row

C1.23

C1.23

C1.23

C1.23

cell calibration

For detailed information refer to calibration on page 30

C1.23.1

temp. comp.

Menu for temperature measurement.

Options:

off: temperature measurement is disabled.

manual: temperature value has to be entered manually.

automatic: temperature measurement is performed as configured

.book Page 26 Friday, January 24, 2014 12:32 PM