Operation – KROHNE OPTISENS COND 1200 EN User Manual
Page 26
4
OPERATION
26
OPTISENS COND 1200
www.krohne.com
01/2014 - 4001967301 - MA OPTISENS COND 1200 R01 en
C1, process input A
C1, process input A
C1, process input A
C1, process input A
C2, process input B
C2, process input B
C2, process input B
C2, process input B
Level
Designation / function
Settings / descriptions
Process input A and B can be either a sensor 1 or a sensor 2. Further information about the type of sensor 1 or 2 please
refer to MAC 100 manual "Sensor input combinations". Process input A is always present, process Input B can be
present.
Note: The exchange of a sensor 1 with a sensor 2, or vice versa, can only be done by the manufacturer!
Depending on the sensor which is connected to a slot A or B the menu changes.
C1.1
C1.1
C1.1
C1.1
parameter (conductivity)
This menu item is for selecting the probe which is connected to process
input A/B. The entries of this selection depends on the chosen device
configuration. The device configuration is customer specific and set during
production.
C1.2
C1.2
C1.2
C1.2
cell constant
Enter cell constant.
C1.3
C1.3
C1.3
C1.3
cable resistance
Enter cable resistance in Ohm
C1.14
C1.14
C1.14
C1.14
time constant
Enter time constant.
C1.15
C1.15
C1.15
C1.15
temperature
Menu for temperature measurement. Available for sensor 1 and sensor 2.
C1.15.1
probe
Options:
•
manual: used if no internal or external temperature sensor is connected to
the signal converter
•
Pt1000: used if an external Pt1000 temperature sensor is connected to the
signal converter
C1.15.2
manual
Only available if C1.15.1 is set to "manual". If you have chosen "manual"
enter temperature.
C1.15.3
correction
Offset correction for temperature measurement. Not available if C1.15.1 is
set to "manual". If you have chosen “Pt 1000” enter the temperature
correction.
C1.15.4
limitation
Measuring ranges for temperature measurement. Enter limitation
C1.15.5
temp. comp.
Menu for editing the temperature compensation parameters for the
measurement.
Options:
•
off
•
table
•
ultra pure water NACl
•
ultra pure water Nao3
•
upw morpholin
•
ultra pure water NH3
•
ultra pure water HCL
•
ultra pure water
•
linear
C1.15.6
temp. coefficient
If you have chosen “linear” enter the temperature coefficient.
C1.15.7
ref. temperature
If you have chosen “linear” enter the reference temperature.
C1.15.8
Reference points
Only available if C1.15.5 is set to “table”. Choose:
•
2 till 10
C1.15.9
tbl. temperature
Only available if C1.15.5 is set to “table”. Choose:
•
temperature for each row
C1.15.10
tbl. conductivity
Only available if C1.15.5 is set to “table”. Choose:
•
conductivity for each row
C1.23
C1.23
C1.23
C1.23
cell calibration
For detailed information refer to calibration on page 30
C1.23.1
temp. comp.
Menu for temperature measurement.
Options:
•
off: temperature measurement is disabled.
•
manual: temperature value has to be entered manually.
•
automatic: temperature measurement is performed as configured
.book Page 26 Friday, January 24, 2014 12:32 PM