4.3.3 Calibration log

Accessing the calibtration log

• You are on the data level and you see the calibration history. With the help of

or

you can

scroll through the different entries.

• If you want to return to the measuring mode press ^

^

^

^ several times until you reach this mode.

INFORMATION!
In order to show the history of the calibrations, the converter has a calibration logbook function.
Up to 64 entries of the calibration history are stored including date and time.

Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup

quick setup

quick setup

quick setup is highlighted.

Press

or

until the main menu test

test

test

test is highlighted.

MAIN MENU

A quick setup

> B test

> B test

> B test

> B test

C setup

D service

Press >>>> to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "test" and "B1" appears,

beneath the submenu sim.process input A

sim.process input A

sim.process input A

sim.process input A is highlighted.

Press or until the submenu logbooks

logbooks

logbooks

logbooks is highlighted.

Press >>>> to enter the chosen menu.

You are on the second submenu level. In the upper line of the display "logbooks" and

"B5.1" appears, beneath the submenu status log

status log

status log

status log is highlighted.

Press

or

until the submenu calibration log

calibration log

calibration log

calibration log is highlighted.

Press >

>>> to enter the chosen menu.