beautypg.com

Installation, 6 calibrating the sensor, 7 installing the sensor – KROHNE OPTISENS COND 1200 EN User Manual

Page 16

background image

3

INSTALLATION

16

OPTISENS COND 1200

www.krohne.com

01/2014 - 4001967301 - MA OPTISENS COND 1200 R01 en

3.6 Calibrating the sensor

Before the sensor is installed, it has to be calibrated. Proceed as described on page 27. Then
continue with the installation procedure.

3.7 Installing the sensor

3.7.1 General installation instructions

WARNING!
Ensure that the pipe is without pressure before installing or removing a sensor!

INFORMATION!
This installation procedure is only recommended for very clean water without any particles in
the water. Otherwise turn the flow-through assembly and be sure that the tube is completely
filled with water, otherwise the measuring reading is wrong.

INFORMATION!
During installation you should fix a shut-off valve in front of and behind the instrument so that
the sensor can be taken out of the bypass in case of check.

INFORMATION!
To achieve reliable measuring results, note the following items:

Always install the sensor in the designated flow through assembly.

The sensor must always have full contact with the measuring medium.

Figure 3-3: Installation requirements for the sensor

1 Flow

min

= 30 l/h / 7.93 gal/h

.book Page 16 Friday, January 24, 2014 12:32 PM