3.6 Calibrating the sensor
Before the sensor is installed, it has to be calibrated. Proceed as described on page 27. Then
continue with the installation procedure.
3.7 Installing the sensor
3.7.1 General installation instructions
WARNING!
Ensure that the pipe is without pressure before installing or removing a sensor!
INFORMATION!
This installation procedure is only recommended for very clean water without any particles in
the water. Otherwise turn the flow-through assembly and be sure that the tube is completely
filled with water, otherwise the measuring reading is wrong.
INFORMATION!
During installation you should fix a shut-off valve in front of and behind the instrument so that
the sensor can be taken out of the bypass in case of check.
INFORMATION!
To achieve reliable measuring results, note the following items:
•
Always install the sensor in the designated flow through assembly.
•
The sensor must always have full contact with the measuring medium.
Figure 3-3: Installation requirements for the sensor
1 Flow
min
= 30 l/h / 7.93 gal/h