• Enter the measured temperature. Press ^ to confirm the entered value. The manually

measured temperature will now be used for temperature compensation.

4.3.2 Calibrating measurement

Calibration is necessary in regular intervals or when installing a new sensor.

When calibrating a conductivity measurement, keep in mind that the measuring system as a
whole is calibrated, and not only the sensor. Therefore the measuring system has to be
recalibrated if, for example, the measuring medium changes.

In an intact sensor, the optimal slope is 59 mV and the optimal zero point is 0 mV. The slope
should at least have a value between 50...65 mV.
Re-calibrate the sensor if the slope does not approximate those limits.
As the sensor ages, the slope gets flatter and the zero error increases. When one or both of
these values exceed certain limits, the signal converter displays a message indicating that the
sensor has to be exchanged.

To avoid alarms on the distrubted control system (DLC) when temporarily removing the sensor
(i.e. for maintenance), the signal converter has a hold function. This function "freezes" all
outputs (i.e. the display and the current outputs) of the last measured value.

Step 2b: probe manual

probe manual

probe manual

probe manual

Measure the temperature of the measuring medium.

Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu setup

setup

setup

setup is highlighted.

Press

or

until the main menu setup

setup

setup

setup is highlighted.

MAIN MENU

A quick setup

B test

> C setup

> C setup

> C setup

> C setup

D service

Press >>>> to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "setup" and "c1" appears,

beneath the submenu process input A

process input A

process input A

process input A is highlighted.

Press

or

to select process input A

process input A

process input A

process input A.

Press >>>> to enter the chosen menu process input A

process input A

process input A

process input A.

You are on the second submenu level. Press or until the submenu temperature

temperature

temperature

temperature is

highlighted.

Press >>>> to enter the chosen menu. The submenu probe

probe

probe

probe is highlighted.

Press

or

until the submenu manual

manual

manual

manual is highlighted.

Press ^ to enter the chosen menu

INFORMATION!
As an indication that the hold function is active, the "warning sign" in the upper left corner of the
display appears. Meanwhile the status messages show "checks in progress". For more details
about how to select the manual hold function refer to the signal converter documentation.