beautypg.com

KROHNE UFM 610P EN User Manual

Page 35

background image

35

Pressing ENTER now will make the instrument search for a
temperature signal. If this is not found then the display will read
the following.

ATTACH SENSORS yy-mm-dd hh:mm:ss

No signal from temp sensor

Press ENTER to try again or

SCROLL to enter a value

The user can now enter a temperature value between -20

°

C and +220

°

C, press ENTER for the separation distance.

The display will now read the following. The temperature will only
be shown on this display if entered manually.

ATTACH SENSORS yy-mm-dd hh:mm:ss

FLUID TEMPERATURE (

°

C)

20.0

Set sensor separation to 33.5

Press ENTER to continue

Now set the transducers to the required separation distance.
Pressing ENTER will take the instrument into flow mode.

READ FLOW yy-mm-dd hh:mm:ss

ERROR MESSAGES APPEAR HERE

Battery
100%
Signal
100%
Temp

+ Total 1564 l

20

°

C

- Total

l

100.0

l/m