18

3.2.1

Attach sensors

The instrument will now provide the user with details on the type of sensor to be attached to the pipe and the mode of operation.
It will also give the approximate maximum flow that can be achieved with the sensors that have been selected.

It is possible to change the flow units at this stage to display the
maximum volumetric flow. Use the keypad to select a flow unit.
Now connect the RED, BLUE and BLACK sensor cables, between
the guide rail and the electronics.

ATTACH SENSORS yy-mm-dd hh:mm:ss

Attach sensor set A in REFLEX mode

(RED connector upstream)

Approx. max. flow:

7.20 m/s

press ENTER to continue

or SCROLL to select another sensor

If the instrument cannot find a temperature signal because the black
sensor cable is not connected, it asks the user to try again. Pressing
ENTER will make the instrument try again or scroll will prompt the
user to enter a value. When a value is entered press ENTER.

ATTACH SENSORS

yy-mm-dd hh:mm:ss

No signal from temp sensor

Press ENTER to try again or

SCROLL to enter a value

Pressing ENTER at this point will give the user the separation
distance or ask for a temperature to be entered.

ATTACH SENSORS yy-mm-dd hh:mm:ss

FLUID TEMPERATURE (

°

C) 20.0

Set sensor separation to

34

Press ENTER to continue

NOTE:
The fluid temperature will only be displayed when entered manually.
The separation distance is displayed in mm.