When the 4mA is adjusted press ENTER. If the 4-20mA is not
connected then the instrument will still display the DAC number
but display Error instead of OK.

CALIBRATE 4-20mA yy-mm-dd hh:mm:ss

Adjust the output current to 4mA

Use UP/DOWN to set, 5/6 to trim

DAC value: 8590 mA OK

Press ENTER when done

Now adjust the 20mA, press ENTER when complete and the
display will return to the SETUP UFM 610 P menu.

CALIBRATE 4-20mA yy-mm-dd hh:mm:ss

Adjust the output current to 20mA

Use UP/DOWN to set, 5/6 to trim

DAC value: 39900 mA OK

Press ENTER when done

If the load is not connected or too high ERROR will be displayed
next to mA, as shown below.

CALIBRATE 4-20mA yy-mm-dd hh:mm:ss

Adjust the output current to 20mA

Use UP/DOWN to set, 5/6 to trim

DAC value: 39900 mA ERROR

Press ENTER when done

3.9.2

Backlight

Use the scroll key to select backlight and press ENTER.

SETUP UFM 610 P yy-mm-dd hh:mm:ss

Set date & time

yy-mm-dd hr-min-sec

Calibrate 4-20mA

Backlight Disabled

Application options
Sensor parameters
Factory settings
Exit menu

This allows the user to enable or disable the backlight. Use the
scroll key to select and press ENTER.

Backlight yy-mm-dd hh:mm:ss

Enabled

Disabled