OPERATION
6
53
WATERFLUX 3070
www.krohne.com
06/2013 - 4001410205 - HB WATERFLUX 3070 R05 en
6.3.11 Battery
After a change of battery:
• Reset the bettery lifetime counter (Menu number B2)
• Select the battery type, if a different type of battery is used. (Menu number B0)
• Change the total battery capacity, if a different type of battery is used. (Menu number B1)
No.
Function
Options
Description
A0
Simulate outputs
0 = Off (default)
If set to 1 (=On) the pulse and status
output automatically generate 1
pulse per second. Set back to 0 to
stop the simulation.
1 = On
A1
Reset all counters
88888 = Off (default)
Set to 00000 to reset all counters.
The value of A1 will automatically
switch back to 88888 after the reset.
00000 = Reset
A2
Verification mode
0 = Off (default)
The display indicates Pxxxxxxx, the
decimal separator position varies
per diameter. Please refer to
Verification mode
on page 54, for
more details.
1 = On
A3
Add-on
communication
0 = Off (default)
Factory setting
1 = On
No.
Function
Options
Description
B0
Battery type
0 = No battery
A wrong setting influences the
battery lifetime calculation.
1 = One internal battery
(default)
2 = Two internal batteries
3 = External battery pack
B1
Total battery capacity
xx.xxx = xx.xxx Ah
(19.000 Ah Default)
Total of all batteries in Ah. After a
change to a different battery type,
change the setting (19 one battery,
38 two batteries, or 78 external
battery)
B2
Reset battery lifetime
counter
0 = Off (default)
Set the value to 1 to reset the battery
lifetime counter. After a reset, the
menu setting automatically goes
back to 0.
1 = Reset