beautypg.com

Operation – KROHNE WATERFLUX 3070 EN User Manual

Page 52

background image

6

OPERATION

52

WATERFLUX 3070

www.krohne.com

06/2013 - 4001410205 - HB WATERFLUX 3070 R05 en

6.3.9 Meter configuration settings

6.3.10 Service and test

INFORMATION!
The meter constant (menu 92), zero offset calibration (menu 93) and zero selection (menu 94)
are for service purposes only.

No.

Function

Options

Description

90

Meter type

0 = Rectangular

(default)

Factory setting

1 = Not used

91

Meter size

Factory setting

(xxx in mm)

By default the meter size is shown in mm. If

menu number 10 (unit of volume for the counter)

is set to 02, 03, 04, 05 or 06 the meter size is

shown in inches. The meter size is also shown

on the name plate and on the display.

92

Meter constant

Factory setting

(xx.xxx)

The meter constant is also shown on the

nameplate and on the display.

93

Zero offset calibration

0 = Off (default)

After a countdown the meter automatically

switches back to the measuring mode and to the

main display. The setting goes back to zero.

Menu number 94 is automatically set to 1.

1 = In situ

determination of

zero point

94

Zero selection

0 = Factory setting

(default)

Factory setting based on factory calibration

1 = Measured value

The measured value is determined in situ if

menu number 93 is set to 1.

CAUTION!
Menu number A0 Simulate outputs is for service purposes only.

WARNING!
Activating a reset of all counters, means that previous values will be lost and
cannot be re-covered.

WARNING!
Activating the verification mode affects the measurements and the counter value during the test
procedure

INFORMATION!
Activation of add-on communication should be ordered with the meter. To use add-on
communication for an existing installation please contact the service department