6.3.7 Status output

6.3.8 Measurement settings

By default the forward flow direction is set from left to right. The red arrow on the measurement
sensor indicates the same direction. The flow direction can be set from right to left in menu
number 80.

No.

Function

Options

Description

40

Status output 1:

Self check

0 = Off (default)

1 = On

41

Status output 1:

Battery pre warning

0 = Off (default)

Active in case of low battery

(remaining capacity < 10%).

1 = On

42

Status output 1:

Battery final warning

0 = Off (default)

Active in case of very low battery

(remaining capacity < 1%).

1 = On

43

Status output 1:

Empty pipe

0 =Off (default)

Active in case of empty pipe

detection.

1 = On

50

Status output 2:

Self check

0 = Off (default)

1 = On

51

Status output 2:

Battery pre warning

0 = Off (default)

Active in case of low battery

(remaining capacity < 10%).

1 = On

52

Status output 2:

Battery final warning

0 = Off (default)

Active in case of very low battery

(remaining capacity < 1%).

1 = On

53

Status output 2:

Empty pipe

0 = Off (default)

Active in case of empty pipe

detection.

1 = On

No.

Function

Options

Description

80

Flow direction

0 = Forward (default)

Sets the flow direction.

1 = Reverse

81

Measurement interval

1 = 1 sec

Sets the interval time between two

measurements.

The default value of 15 sec is chosen

for an optimal energy consumption

of the battery.

The 1 sec option is for test puposes

only

5 = 5 sec

10 = 10 sec

15 = 15 sec (default)

20 = 20 sec

82

Low flow cut off

0 = 0 mm/s

Low flow cut off in mm/s.

Measurements below this value are

neglected.

5 = 5 mm/s

10 = 10 mm/s (default)

83

Empty pipe detection

(EP)

0 = Off (default)

If switched to 1 (=on), the display

shows "- EP -" in case of empty pipe

and the measurement is stopped.

In case of verification to MID Annex

MI-001, the EP is set to on.

1 = On