6.3.7 Status output
6.3.8 Measurement settings
By default the forward flow direction is set from left to right. The red arrow on the measurement
sensor indicates the same direction. The flow direction can be set from right to left in menu
number 80.
No.
Function
Options
Description
40
Status output 1:
Self check
0 = Off (default)
1 = On
41
Status output 1:
Battery pre warning
0 = Off (default)
Active in case of low battery
(remaining capacity < 10%).
1 = On
42
Status output 1:
Battery final warning
0 = Off (default)
Active in case of very low battery
(remaining capacity < 1%).
1 = On
43
Status output 1:
Empty pipe
0 =Off (default)
Active in case of empty pipe
detection.
1 = On
50
Status output 2:
Self check
0 = Off (default)
1 = On
51
Status output 2:
Battery pre warning
0 = Off (default)
Active in case of low battery
(remaining capacity < 10%).
1 = On
52
Status output 2:
Battery final warning
0 = Off (default)
Active in case of very low battery
(remaining capacity < 1%).
1 = On
53
Status output 2:
Empty pipe
0 = Off (default)
Active in case of empty pipe
detection.
1 = On
No.
Function
Options
Description
80
Flow direction
0 = Forward (default)
Sets the flow direction.
1 = Reverse
81
Measurement interval
1 = 1 sec
Sets the interval time between two
measurements.
The default value of 15 sec is chosen
for an optimal energy consumption
of the battery.
The 1 sec option is for test puposes
only
5 = 5 sec
10 = 10 sec
15 = 15 sec (default)
20 = 20 sec
82
Low flow cut off
0 = 0 mm/s
Low flow cut off in mm/s.
Measurements below this value are
neglected.
5 = 5 mm/s
10 = 10 mm/s (default)
83
Empty pipe detection
(EP)
0 = Off (default)
If switched to 1 (=on), the display
shows "- EP -" in case of empty pipe
and the measurement is stopped.
In case of verification to MID Annex
MI-001, the EP is set to on.
1 = On