Operation – KROHNE WATERFLUX 3070 EN User Manual

Page 44

background image

6

OPERATION

44

WATERFLUX 3070

www.krohne.com

06/2013 - 4001410205 - HB WATERFLUX 3070 R05 en

Status output

Status output

Status output

Status output

40

Status output 1:

self-check

0

Off

41

Status output 1:

battery pre warning

0

Off

42

Status output 1:

battery final warning

0

Off

43

Status output 1:

empty pipe

0

Off

50

Status output 2:

self-check

0

Off

51

Status output 2:

battery pre warning

0

Off

52

Status output 2:

battery final warning

0

Off

53

Status output 2:

empty pipe

0

Off

Measurement settings

Measurement settings

Measurement settings

Measurement settings

80

Flow direction

0

Forward

81

Measurement interval

15

15 sec

82

Low flow cut off

10

10 mm/s

83

Empty pipe detection (EP)

0

Off

Meter configuration

Meter configuration

Meter configuration

Meter configuration

90

Meter type

0

Factory setting

91

Meter size

100

Factory setting (example)

92

Meter constant

01.000

Factory setting (example)

93

Zero offset calibration

0

Off

94

Zero selection

0

Factory setting (calibration settings)

Service and test

Service and test

Service and test

Service and test

A0

Simulate outputs

0

Off

A1

Reset all counters

88888

Off

A2

Flow verification on site

0

Off

A3

Add-on communication

0

Factory setting

Battery

Battery

Battery

Battery

B0

Battery type

1

Internal single battery

B1

Battery capacity

19.000

19 Ah

B2

Reset battery lifetime counter

0

Off