3-24

BOP HIPWR 101509

NOTE: If a value to be saved is beyond the range of the unit, the value saved will be limited to

the maximum value. If the Mode is changed, the numerical values for the main and
protection channels must be revised accordingly.

• To modify the parameter, press

!

. Highlight the desired choice from a list, or use the

keypad to enter a numeric value, then press

$

to save, or

%

to abort.

• To exit without changing the parameter, press

%

.

5. Continue to modify parameters as described in step 4 above. When all parameters have

been modified, press

$

to save for power-up. Press

%

to exit without storing the

changes.

NOTE: If the message CHANGES NOT ALLOWED is displayed, the memory location has been

remotely locked (see PAR. A.12 for details).

6. After the changes have been saved, the unit returns to the Saved Settings screen, where

the saved settings can now be applied to the output by pressing

@

.

TABLE 3-8. SAVE/RECALL MENU

PARAMETER

CHOICES

(BOLD = Factory Default)

FUNCTION

NOTE: DEFAULT values for empty cells are the settings of the unit at the time the save/recall menu is entered.

MODE

VOLT

CURRENT

EXTERNAL

VOLT - Selects voltage mode. Voltage value determined by SETTING.
CURRENT - Selects current mode. Current value determined by SETTING.
EXTERNAL - Used for customized configuration where EXT VM/CM signal at pin
2 of I/O port controls mode; consult factory for details.

MAIN CHANNEL

REFERENCE

INTERNAL

EXTERNAL

EXTL REF LVL

INTERNAL: Any external reference applied to the Analog I/O port pin 11 is ignored
and the internal reference established by SETTING field.
EXTERNAL: The internal reference is ignored, and the external reference applied
to the Analog I/O port pin 11 is used (see PAR. 3.4.3.1).
EXTL REF LVL: Allows the gain of the BOP to be established by the maximum
output voltage or current introduced by the SETTING field. (see PAR. 3.4.3.2).

SETTING

nn.nnn or

nnn.nnn

(model and parameter

dependent)

Sets the active channel setting. E.g., if VOLT mode is selected, nnn.nnn is voltage
setting, if CURRENT mode is selected, nn.nnn is current setting. If EXTERNAL
mode is selected, SETTING is not available.

PROTECTION

MODE

INTERNAL

EXTERNAL

LESSER LIMIT

INTERNAL - Allows limits to be controlled by POSITIVE and NEGATIVE values.
EXTERNAL: Allows limits to be controlled by analog signals applied to the analog
port (see PAR. 3.4.4).
LESSER LIMIT: Allows protect limit to be automatically selected from either 1) the
external analog voltage applied to the Analog I/O port or 2) the value set in the
POSITIVE and NEGATIVE fields. Whichever limit has a lower absolute value
(closest to zero) has effect (see PAR. 3.4.4).

POSITIVE nn.nnn

or

nnn.nnn

This is the positive value for Voltage protect if CURRENT mode is selected, or the
positive value for Current protect if VOLTAGE mode is selected. If EXTERNAL
mode is selected, this field is not accessible.

NEGATIVE

nn.nnn or

nnn.nnn

This is the negative value for Voltage protect if CURRENT mode is selected, or the
negative value for Current protect if VOLTAGE mode is selected. If EXTERNAL
mode is selected, this field is not accessible.

OUTPUT

OFF

ON

OFF - Output disabled. The behavior of the unit when output is disabled depends
on the LOAD TYPE setting (see PAR.3.3.6).
ON - Output enabled.