4-4
BOP-1K 101509
4.3
CALIBRATION USING REMOTE SCPI COMMANDS VIA GPIB OR RS 232 INTERFACE
The primary means of adjustment are the CAL:DATA commands. Basically these commands
are fine and coarse adjustments, however there are slight differences in how they work for dif-
ferent types of calibrations.
a. ZERO Calibrations (always done first). The only means of adjustment is the
CAL:DATA
ing at 0, with 2047 increments in either direction to achieve maximum output levels of ±2.5% of
E
Onom
. Adjust the output to obtain the closest value to zero.
b. MAX and MIN Calibrations. The primary means of adjustment is the CAL:DATA
command to either increase (CAL:DATA 1) or decrease (CAL:DATA -1) the output value.
Adjustment can be done in larger increments by increasing the numeric value, e.g.,
CAL:DATA +10 or CAL:DATA -10. Adjust the output to obtain the closest value above the full
scale value.
c. VPR and CPR Calibrations. The only means of adjustment is the CAL:DATA
command which provides a total of 4095 increments of adjustment from zero to maximum.
Adjust the output to obtain the closest value above the full scale value.
TABLE 4-3. VOLTAGE CALIBRATION MEASUREMENTS AND TOLERANCES
MODEL
VOLTAGE
ZERO
±FULL SCALE
VOLTAGE
±FULL SCALE
VPR LIMIT
BOP 10-75MG
0V
±0.0001V
10V (MAX)
-10V (MIN)
±0.001V
10V (MAX)
-10V (MIN)
±0.003V
BOP 20-50MG
0V
±0.0002V
20V (MAX)
-20V (MIN)
±0.002V
20V (MAX)
-20V (MIN)
±0.006V
BOP 36-28MG
0V
±0.0036V
36V (MAX)
-36V (MIN)
±0.0036V
36V (MAX)
-36V (MIN)
±0.01V
BOP 50-20MG
0V
±0.005V
50V (MAX)
-50V (MIN)
±0.005V
50V (MAX)
-50V (MIN)
±0.015V
BOP 72-14MG
0V
±0.0072V
72V (MAX)
-72V (MIN)
±0.0072V
72V (MAX)
-72V (MIN)
±0.022V
BOP 100-10MG
0V
±0.01V
100V (MAX)
-100V (MIN)
±0.01V
100V (MAX)
-100V (MIN)
±0.03V