B.14 OUTPut[:STATe]

QUERY

OUTP?

Syntax:

Short Form: OUTP[:STAT]?

Long Form: OUTPut[:STATe]?

Return Value: (0 or 1)

Description: Indicates whether power supply output is enabled or disabled. Returns 0 if output disabled,

returns 1 if output enabled. Related Commands: OUTP.

B.15 OUTPut:MODE COMMAND

OUTP:MODE

Syntax:

Short Form: OUTP:MODE {ACTIVE | RESISTIVE | BATTERY}
Long Form: OUTPut:MODE {ACTIVE | RESISTIVE | BATTERY}

Description: Determines how the unit functions when output is turned off depending on the type of load in

use: active, resistive or battery. See PAR. 3.3.6 for description of ACTIVE, RESISTIVE and BAT-
TERY modes.

B.16 OUTPut:MODE?

QUERY

OUTP:MODE?

Syntax:

Short Form: OUTP:MODE?

Long Form: OUTPut:MODE]?

Return Value: ACTIVE | RESISTIVE | BATTERY

Description: Indicates whether power supply is configured for ACTIVE, RESISTIVE or BATTERY type loads.

See PAR. 3.3.6 for description of ACTIVE, RESISTIVE and BATTERY modes. Related Commands:
OUTP:MODE.

B.17 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] COMMAND

CURR

Syntax:

Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]
= digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Sets programmed current level at power supply output. This command programs output current to

a specific value; actual output current will depend on load conditions. Any data that exceeds I

MAX

causes error message -222 “data out range” to be posted to the error queue (See example, Figure B-
1.)

B.18 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] QUERY

CURR?

Syntax:

Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]? MIN, MAX
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]? MIN, MAX
Return Value: = digits with decimal point and Exponent, e.g., 2.71E1 for 27.1

Description: Returns either the programmed value, maximum value, or minimum value of current. The

CURR? query returns the programmed value of current. Actual output current will depend on load con-
ditions. The CURR?MAX query returns the maximum positive current allowed for a particular model.
CURR? MIN returns minimum negative current allowed for power supply (always negative). Related
Commands:
CURR. (See example, Figure B-1.)

B.19 [SOURce:]CURRent[:LEVel]:LIMit[:BOTH] COMMAND

CURR:LIM[:BOTH]

Syntax:

Short Form: [SOUR:]CURR[:LEV]:LIM[:BOTH]
Long Form: [SOURce:]CURRent[:LEVel]:LIMit[:BOTH]
where is between zero and I

NOM

Description: Establishes the software limit for output current, i.e., sets the maximum value of output cur-

rent that the unit will be allowed to source and sink. Requires MEM:UPD (PAR. B.12) to save for
power up. See PAR. 3.5.2.2 and Figure 3-16 for special programming considerations. (See Figure B-
4)

B.20 [SOURce:]CURRent[:LEVel]:LIMit[:BOTH]? QUERY

CURR:LIM[:BOTH]?

Syntax:

Short Form: [SOUR:]CURR[:LEV]:LIM[:BOTH]?
Long Form: [SOURce:]CURRent[:LEVel]:LIMit[:BOTH]?
Returns ,

Description: Identifies the positive software limit, followed by the negative software limit value of output current that

the unit can source or sink.