4-2

BOP-1K 101509

Remote Calibr.
VOLTAGE
EXTERNAL
PROTECTION,
Pos. Voltage
Neg. Voltage
(together)

1. ZERO
2. POSITIVE

External:
1. 0.0V
2. +10.0V
(Analog I/O Port
pin 14+6 ref to
pin12)

N/A

No requirements

1. Automatic internal
measurement
2. Remote Calibration
only.

CURRENT,
INTERNAL

1. ZERO
2. POSITIVE
3. NEGATIVE

Internal

Output Current:
1. 0.0A
2. +Current FS
3. –Current FS

Short with Meas.
Shunt
Automatic CM

1. DVM between probe
terminals of Meas.
Shunt (LO towards
COM S)
2. Adjust for zero, posi-
tive and negative FS

CURRENT,
PROTECTION

1. POSITIVE
2. NEGATIVE

Internal

Output Current:
1. +Current FS
2. -Current FS

Short with Meas.
Shunt
Automatic VM

1. DVM between probe
terminals of Meas.
Shunt (LO towards
COM S)
2. Adjust for positive
and negative FS.

CURRENT,
EXTERNAL

1. ZERO
2. POSITIVE

External:
1. 0.0V
2. +10.0V
(Analog I/O Port
pin 11 ref to 4)

Output Current:
1. 0.0A
2. +Current FS

Short with Meas.
Shunt
Automatic CM

1. DVM between probe
terminals of Meas.
Shunt (LO towards
COM S).
2)Adjust for zero, pos
FS.

CURRENT,
GAIN

1. ZERO
2. POSITIVE

External:
1. 0.0V
2. +10.0V
(Analog I/O Port
pin 11 ref to 4)

Output Current:
1. 0.0A
2. 0.0A

Short with Meas.
Shunt
Automatic CM

1. DVM between probe
terminals of Meas.
Shunt (LO towards
COM S)
2. Adjust for zero
(both Zero and Positive)

EXTERNAL
CONTROLS,
PARALLEL

1. ZERO
2. POSITIVE

External:
1. 0.0V
2. +10.0V
(Par/Ser CTRL-IN
Port pin3 ref. to
pin1)

Output Current:
1. 0.0A
2. +Current FS

Short with Meas.
Shunt
Automatic CM

1. DVM between probe
terminals of Meas.
Shunt (LO towards
COM S)
2. Adjust for zero, pos.
FS
3. Calibrate unit for
Slave in Parallel

EXTERNAL
CONTROLS,
I_OUT

1. ZERO
2. POSITIVE

INTERNAL

Voltage at Par/Ser
CTRL-IN Port pin 6
ref. to pin1:
1. 0.0V
2. +10.0V

Short with Meas.
Shunt
Automatic CM

1. DVM at Par/Ser
CTRL-IN Port pin 6 ref.
to pin 1
2. Adjust for zero,
+10.0V
3. Calibrate unit for
Master in Parallel

Remote Calibr.
CURRENT
EXTERNAL
PROTECTION,
Pos. Current
Neg. Current
(together)

1. ZERO
2. POSITIVE

External:
1. 0.0V
2. +10.0V
(Analog I/O Port
pins 13, 5 ref to
pin 12)

N/A

No requirements

1. Automatic internal
measurement
2. Remote Calibration.
only.

TABLE 4-1. CALIBRATION SUMMARY (CONTINUED)

Type

Step

Reference Type

and Value

Monitored
Parameter

Output Conditions

Notes