B.71 [source:]list:set:trigger? query, B.72 [source:]list:set:wait command, B.73 [source:]list:set:wait? query – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.2.48 to 3.04 User Manual
Page 169: B.74 [source:]list:trigger command, B.75 [source:]list:voltage command, B.76 [source:]list:voltage? query, B.71, Sour, B.72, B.73
BOP-1K 101509
B-19
B.71 [SOURce:]LIST:SET:TRIGger? QUERY
LIST:SET:TRIG?
Syntax:
Short Form: LIST:SET:TRIG?
Long Form: LIST:SET:TRIGger?
Returns
Description: Returns trigger pulse-width. If trigger pulse duration has not been set, error 100- “command error”
results
.
B.72 [SOURce:]LIST:SET:WAIT COMMAND
LIST:SET:WAIT
Syntax:
Short Form: LIST:SET:WAIT value
Long Form: LIST:SET:WAIT value
where:
Description: Establishes the time-out for a wait command. If this command is not sent, the time out period will
be 0 indicating an infinite wait time. (See Figures B-6 and B-7.)
B.73 [SOURce:]LIST:SET:WAIT? QUERY
LIST:SET:WAIT?
Syntax:
Short Form: LIST:SET:WAIT?
Long Form: LIST:SET:WAIT?
Returns
Description: Returns the time-out for a wait command. Returns 0 if LIST:SET:WAIT not sent, indicating infinite wait
time.
B.74 [SOURce:]LIST:TRIGger COMMAND
LIST:TRIG
Syntax:
Short Form: LIST:TRIG
Long Form: LIST:TRIGger
where:
Description: Similar to LIST:VOLT or LIST:CURR, except also allows Output Off flag of External Protection
Port to be used to trigger an external device. In addition to setting the output to the
fied (for duration established by LIST:SET:TRIG), causes the Output Off flag transistor (EXT_E, pin 5,
and EXT_C, pin 6, of the External Protection port) to be set to the state (on or off) for the duration set
by the LIST:SET:TRIG command.
If the trigger duration is not established by the LIST:SET:TRIG command, a -221 “Settings conflict”
error results. This command must be preceded by LIST:CURR or LIST:VOLT command, otherwise a -
221 “Settings conflict” error results. The entries on the dwell list must match the values on the current
or voltage list, otherwise a -226 “Lists Not Same Length” error results. (See Figure B-7.)
B.75 [SOURce:]LIST:VOLTage COMMAND
LIST:VOLT
Short Form: LIST:VOLT[:LEV]
Long Form: LIST:VOLTage[:LEVel]
Description: Adds the voltage value (in Amps) to list. This command sequentially adds LIST:VOLTage values to
the main channel List Data Table locations illustrated in Table B-2. LIST:CLE sets starting location to 0.
Next location indicated by LIST:VOLT:POIN? The maximum number of entries is 2950, 3933, or 5900.
Since the input buffer of the BOP has a limit of 253 characters, multiple commands are necessary to
complete the full 5900 entries of the list. If LIST:CURR has any entries, an error message: -221,"Set-
tings conflict" is posted in the error queue. Related Commands: LIST:VOLT:POIN?, FUNC:MODE,
LIST:CLE, *RST. (See example, Figure B-5.)
B.76 [SOURce:]LIST:VOLTage? QUERY
LIST:VOLT?
Syntax:
Short Form: LIST:VOLT?
Long Form: LIST:VOLTage?
Return Value:
Description: Identifies the parameters (main channel) entered for the list. Starting at location established by
LIST:QUERy, returns comma-separated list of up to 16 values indicating the main channel parameters
entered. i.e., the contents of the main channel locations of Table B-2. Related Commands: LIST:
VOLT, LIST:QUERy. If LIST:CURR has any entries, an error message: -221,"Settings conflict" is
posted in the error queue. (See example, Figure B-5.)