3.5.3.1.3 DETERMINING WHETHER *RST COMMAND SETS THE OUTPUT OFF OR ON

The user can control whether the *RST command (see PAR. A.11) sets the output on or off.
Table 3-17 describes how the unit operates depending on how *RST set Output is config-

ured, and the load type selected.

From the power-up screen press

%

, and note whether “*RST set Output” setting is as

desired. To modify highlight Interface Settings (see Table 3-16) and press

!

, then highlight On

or Off and press

!

to save. Press

$

to save for power-up or

%

to apply the changes (with-

out saving for power-up) and exit.

3.5.3.2

RS 232 SERIAL PORT SETUP

Paragraphs 3.5.3.2.1 through 3.5.3.2.3 below describe the setup required for the BOP to com-
municate via the RS 232C Serial bus using SCPI commands.

3.5.3.2.1 SELECT BAUD RATE

To change the Baud Rate configuration press

%

from the power-up screen, then press

!

to

enter the Interface menu (see Table 3-16). Highlight Baud Rate and press

!

, then highlight the

Off (to disable the serial port), 9600 or 19,200 and press

$

to save. Press

$

to save for

power-up or

%

to apply the changes (without saving for power-up) and exit.

3.5.3.2.2 CONFIGURE XON/XOFF PROTOCOL

From the power-up screen press

%

, then press

!

to enter the Interface menu (see Table 3-

16), highlight SERIAL XON/XOFF and press

!

. Highlight the desired option and press

$

to

save, then press

$

to save for power-up or

%

to apply the changes (without saving for

power-up) and exit.
Enable - Allows BOP to control when data is received (see PAR. 3.5.6.2.1 for a description of
Xon/Xoff protocol).
Disable - Disables XON/XOFF. If both Serial XON/XOFF and Prompt functions are disabled,
Echo mode is set (see PAR. 3.5.6.2.2 for a description of echo mode).

3.5.3.2.3 CONFIGURE PROMPT MODE

From the power-up screen press

%

, then press

!

to enter the Interface menu (see Table 3-

16). Highlight Serial Prompt and press

!

, highlight the desired option and press

$

to save,

TABLE 3-17. OPERATION OF #RST COMMAND

Load type when *RST

Issued (see NOTE)

BOP Status After *RST Issued

Internal

Front Panel Display

• Active or Resistive Load

(see PAR. 3.3.6)

• Voltage mode
• Main Channel reference set to 0.0V
• Current protect set to 5% of

I

Onom

• Voltage protect set to 5% above

E

Onom

• LCD shows mode is Voltage, Source
• Vset set to 0.0V
• ±CProt set to minimum (box) values
• If *RST set to OFF, output is disabled
• If *RST set to ON, output is enabled

• Battery Load (see PAR.

3.3.6)

• If *RST set to ON, unit is set to Voltage Mode
• If *RST set to OFF unit is set to Current Mode
• Main Channel reference set to 0.0A
• Current protect set to 5% of

I

Onom

• Voltage protect set to 5% above

E

Onom

• LCD shows mode is Voltage, Source
• Vset set to 0.0V
• ±CProt: minimum (box) values
• If *RST set to OFF, output is disabled
• If *RST set to ON, output is enabled

• NOTE: Prior to issuing *RST unit may be operating in either voltage mode or current mode.