3.11

CURRENT READING CALIBRATION (R20)

NOTE: The Full Scale Output Current Adjustment (PAR. 3.9) must be done before performing

the following procedure, and the PRECISION 4-TERMINAL SHUNT should be still
connected to the power supply output terminals.

1. Program the BOP Power Supply for PLUS FULL SCALE CURRENT, less one percent (verify

by reading external DVM). E.g., for 20V

• If using SCPI send

FUNC:MODE CURR;:CURR zz;VOLT MAX

• If using CIIL send

FNC DCS:CH1 CURR zz with VLTL 20

where

zz = aa - (aa/100)

where aa is the maximum rated output current of the

BOP (e.g., for BOP 20-10M,

zz

= 10-(10/100) = 9.9).

2. MEASURE the output current using programming commands, then check the readback

data.

• In SCPI send

MEAS:CURR?

then check the readback data.

• In CIIL: send

FNC DCS:CH1 GAL

F0

INX CURR

FTH CURR

NOTE: For subsequent measurements using CIIL it is only necessary to send

INX

CURR

and

FTH CURR

commands.

3. Locate CURRENT READ. CAL. control R20 (see Figures 2-1 and 3-1, refer to Table 3-1).

4. Adjust control R20, until the measured value matches the programmed value using MEA-

SURE commands (see step 2) as needed.