beautypg.com

6 voltage reading zero calibration (r35), Voltage reading zero calibration (r35) -3, Bop power supply, internal calibration controls -3 – KEPCO BIT 4882F User Manual

Page 23

background image

BIT 4882 030507

3-3

2. MEASURE the output voltage at the sense terminals at the BOP front panel using the DVM.

3. Locate VOLTAGE FULL SCALE control R21 (see Figures 2-1 and 3-1, refer to Table 3-1).

4. Adjust control R21 as needed until DVM reads FULL SCALE VOLTAGE, ±1 millivolt.

3.6

VOLTAGE READING ZERO CALIBRATION (R35)

1. Program the BOP power supply for ZERO VOLTAGE and MAXIMUM CURRENT LIMIT.

• If using SCPI send

VOLT 0;CURR MAX

• If using CIIL send

FNC DCS:CH1 VOLT 0 with CURL .5

TABLE 3-1. BOP POWER SUPPLY, INTERNAL CALIBRATION CONTROLS

REFERENCE

DESIGNATION

CONTROL

NAME

PURPOSE

ADJUSTMENT

PROCEDURE

(PAR.)

R19 *

VOLTAGE READING

‘MEASURE’d voltage reading adjustment

3.7

R20 *

CURRENT READING

‘MEASURE’d current reading adjustment

3.11

R21 *

VOLTAGE FULL SCALE

Full scale output voltage adjustment

3.5

R22 *

CURRENT FULL SCALE

Full scale output current adjustment

3.9

A1R31, A1R32

( ± ) 10V CAL.

Reference voltage calibration

3.2

R35 *

VOLTAGE READBACK ZERO

Zero output voltage readback adjustment

3.6

R36 *

CURRENT READBACK ZERO

Zero output current readback adjustment

3.10

R50 *

AMMETER ZERO

Sensing amplifier offset adjustment

3.3

A1R81

EO COMP AMP ZERO

Voltage channel zero adjustment

3.4

A1R83

IO COMP AMP ZERO

Current channel zero adjustment

3.8

* Located on BIT 4882 card

This manual is related to the following products: