beautypg.com

Figure b-11. voltage:triggered command, Figure b-12. voltage:triggered query, B-11 vo – KEPCO BIT 232F User Manual

Page 45: B-12 vo, Volt:trig

background image

BIT 232 022800

B-7

FIGURE B-11. VOLTage:TRIGgered COMMAND

FIGURE B-12. VOLTage:TRIGgered QUERY

Syntax:

Short Form:

VOLT:TRIG

Long Form:

[SOURce]:VOLTage:TRIGgered

Function:

Programs voltage value to be implemented by *TRG command. Actual output voltage will depend on
load conditions.

Description:

This command can be used to reset many power supplies to preselected parameters by issuing a sin-
gle *TRG command.

Example:

VOLT 21; CURR 1.1

Power supply commanded to go to 21V, 1.1A

VOLT:TRIG 29.3

Power supply current programmed to 29.3 upon receipt of *TRG

VOLT?

Controller reads 21, indicating programmed current value = 21V

*TRG

Power supply commanded to voltage value established by VOLT:TRIG
command (29.3) and current value established by CURR:TRIG

VOLT?

Controller reads 29.3, indicating programmed current value = 2.3A

VOLT:TRIG

Syntax:

Short Form:

VOLT:TRIG?

Long Form:

[SOURce]:VOLTage:TRIGgered?

Function:

Returns value to controller representing voltage value to be programmed by *TRG command.

Description:

This command returns the voltage value established by VOLT:TRIG command.

Example:

VOLT 21; CURR 1.1

Power supply commanded to go to 21V, 1.1A

VOLT:TRIG 29.3

Power supply voltage programmed to 29.3 upon receipt of *TRG

VOLT?

Controller reads 21, indicating programmed voltage value = 21V

VOLT:TRIG?

Controller reads 29.3 (voltage value established by VOLT:TRIG
command

VOLT:TRIG?