Technical specifications – HT instruments PVCHECK User Manual
Page 61
![background image](/manuals/367529/61/background.png)
PVCHECK
EN - 59
10. TECHNICAL SPECIFICATIONS
10.1. TECHNICAL SPECIFICATIONS FOR PV INSTALLATION EFFICIENCY
Accuracy is indicated as
[%reading + (no. of digits) * resolution] at 23°C ± 5°C, <80%HR
DC voltage
Range [V]
Resolution [V]
Accuracy
5.0
199.9
0.1
(1.0rdg + 2dgt)
200.0
999.9
0.5
DC current (by means of external clamp transducer)
Range [mV]
Resolution [mV]
Accuracy
-1100
-5
0.1
(0.5rdg + 0.6mV)
5
1100
The value of current is ALWAYS displayed with a positive sign: the value of current transduced in a voltage lower than 5mV is zeroed.
FS DC clamp [A]
Resolution [A]
Minimum read value [A]
1< FS
10
0.001 0.05
10< FS
100
0.01 0.5
100< FS
1000
0.1 5
DC power (Vmis > 150V)
Clamp FS [A]
Range [W]
Resolution [W]
Accuracy
1< FS
10
0.000k
9.999k
0.001k
(1.5rdg + 3dgt)
(Imis < 10%FS)
(1.5rdg)
(Imis
10%FS)
10< FS
100
0.00k
99.99k
0.01k
100< FS
1000
0.0k
999.9k
0.1k
Vmis = voltage at which power is measured; Imis = measured current
Irradiance (with HT304N reference cell)
Range [mV]
Resolution [mV]
Accuracy
1
40.0
0.02
(1.0rdg + 0.1mV)
Temperature (with probe type PT300N)
Range [°C]
Resolution [°C]
Accuracy
-20.0
100.0
0.1
(1.0rdg + 1°C)