HT instruments PVCHECK User Manual
Page 52

PVCHECK
EN - 50
6.6.
LIST OF DISPLAY MESSAGES
MESSAGE DESCRIPTION
Function not available
The selected function/characteristic is not available
Data not saved
The instrument was not able to save the data
Wrong date
Set a correct system date
RADIO transmission error
The instrument does not communicate via RF with external units
SOLAR-02: Firmware incorrect
Inconsistent SOLAR-02 FW. Update firmware.
Firmware incorrect
Unsuitable instrument FW. Update firmware.
Error 4: contact service
Internal instrument error
Database full
The number of modules added to the internal DB is > 30
Module already added
The name of the module has already been added to the DB
Memory full
Instrument memory full upon pressing the GO key
Error: Vmpp >= Voc
Check module settings inside DB
Error: Impp >= Isc
Check module settings inside DB
Error: Vmpp * Impp >= Pmax
Check module settings inside DB
Error: Alpha too high
Check module settings inside DB
Error: Beta too high
Check module settings inside DB
Error: Gamma too high
Check module settings inside DB
Error: Toll too high
Check module settings inside DB
Wait data analysis
Download data from SOLAR-02 and wait for the result of PV efficiency
test
Download error
Contact service department
Error while saving
Problems with the access to memory area
Remote unit undetected
The instrument does not detect any remote unit SOLAR-02
Unable to perform analysis
Problems with the data downloaded from SOLAR-02. Check settings.
Data unavailable
Generic error. Repeat the test.
Negative voltage
Check the polarity of the instrument’s input terminals
Low voltage
Check the voltage between input terminals P and N
Vin > 1000
Voltage between input terminals > 1000V
Wrong no. of modules. Continue?
Set no. of modules inconsistent with measured Voc
Ref. cell temp. exceeding limits
Temperature measured by reference cell too high
Cell temp. undetected
(ENTER/ESC)
Measurement not carried out on module cell
Low battery
Low battery charge. Insert new batteries.
Please wait for cooling…
Instrument overheated. Wait before resuming the test.
Irradiance too low
Irradiance value lower than the minimum threshold set
NTC Error
Internal NTC efficiency jeopardized. Contact service department
Isc current too high
Measured Isc current > 10A
Current < Lim
Current between P and N lower than the minimum detectable value
EEPROM error: contact service
Internal instrument error
FRAM error: contact service
Internal instrument error
RTC Error: contact service
Internal instrument error
RADIO error: contact service
Internal instrument error
FLASH error: contact service
Internal instrument error
IO EXP error: contact service
Internal instrument error
Voltage > limit
Voltage between terminals E and C > 10V
Label already assigned
Change the numeric reference of the marker associated with the measure
Isc current < Lim
Isc current lower than the minimum detectable value. Contact service
department
CAUTION: internal short-circuit
Contact service department
CAUTION: blown fuse
Contact service department
Reset calibration. Press ENTER.
Resistance value of input cables > 2
Calibration not OK
Calibrated resistance value > measured resistance
Error: Isc offset measurement
Internal instrument error
Rcal > measured R
Calibrated resistance value > measured resistance
CAUTION: AC voltage at P-N
terminals
Presence of AC input voltage
Wait for condenser discharge
Wait for discharge of tested object after insulation