PVCHECK
EN - 19
modules’ Voc.
“Manual”
setting, made by the operator, of the known
temperature value of the module in the line below.
Aux
temperature measurement with auxiliary probe
PT1000.
Tol. Voc (%) percentage value of the desired limit
tolerance (set by the operator according to his/her own
requirements) for the measurement of Voc carried out
by the instrument. Allowable values: 0%
25%. The
value between brackets (4%) indicates the reading
error of the instrument in Voc measurement.
Tol. Isc (%) percentage value of the desired limit
tolerance (set by the operator according to his/her own
requirements) for the measurement of Isc carried out
by the instrument. Allowable values: 0%
25%. The
value between brackets (4%) indicates the reading
error of the instrument in Isc measurement.
Ins Test enabling/disabling of insulation
measurement and setting of test voltage. Possible
options: OFF, 250V, 500V, 1000V. With enabled
function, the line “Ri min” appears, which allows
setting the minimum limit threshold in the range 0.1
100M
RPE Test enabling/disabling of continuity test and
setting of the limit threshold value for measurement.
Possible options OFF, 1
5 in steps of 1
5. Press the SAVE key to save the settings made; the
message “Data saved” will be displayed for a few
seconds. Press the ESC/MENU key to exit without saving
and go back to the previous screen.
CAUTION
The settings saved for test voltage for the insulation measurement carried out
within the IVCK function also affect the settings for the single measurement
(MENU
M
).