HT instruments FULLTEST3 User Manual
Page 9

FULLTEST
3
EN - 8
Clamp current
-
Measurement in combination with HT96U current clamp.
-
Three ranges 1 A, 100 A and 1000 A.
-
Limit value adjustable, visual and acoustic warning in case of exceeded
value.
Leakage current
-
Measurement of I
PE
current on schuko socket (differential method).
-
Measurement with current clamp type HT96U, three ranges 1 A, 100 A
and 1000 A.
-
Limit value adjustable, visual and acoustic warning in case of exceeded
value.
GENERAL advantages
-
Portable Machinery & Panel Tester constructed according to the
IEC/EN60204-1 and IEC/EN61439-1 standards.
-
Operation system WINDOWS EMBEDDED COMPACT 7 supports all
measurements and operations.
-
Easy and clear operation by using the touch screen and intuitive hard
keys.
-
TRMS measurements.
-
Data memory for 999 measurement results, three levels (e.g.
CUSTOMER, LOCATION, MACHINE) plus additional COMMENT.
-
Real time clock included.
-
Integrated interface (USB 2.0) for transfer of measurement results to PC.
-
Separate interface (USB 2.0) for connection of USB barcode reader, USB
keyboard, USB memory stick, printer or IMP57-impedance tester.
-
Graphic colour touch screen 102×60 mm, 480×272 dots.
-
Compact housing with external accessory bag.
-
Quick connection diagrams and limit values under the instrument cover.
-
Fuse protection in case of overload.
-
PC Software TOP VIEW available.
-
Complete test accessories included.
-
Blue tooth communication.
-
Remote START/STOP and SAVE function.
-
*HELP screens available.
-
*20 programmable AUTO test procedures.
* Special FW version.
2.2. OPENING THE INSTRUMENT'S COVER
The instrument is built in a robust plastic case that allows comfortable transport.
We recommend the user to follow the next opening instructions:
1
2
Figure 1: Opening the instrument’s cover
Place the unit on to a hard horizontal surface.
Press on the case cover with your hands, see the
mark 1.
Unlock the fixing hooks of the cover, see the mark 2.
Open the cover into vertical position.