CAUTION
Max external voltage between two R
PE
or between two SENSE test terminals
is 10 VAC, no DC external voltage is allowed! In case of higher external
voltage fuse F4 (
T20A/500V, 6.3×32 mm
) may blow!
Measurement time in MANUAL mode is limited to 60 min!
The following specific information can be shown on the display during measurement:
Information displayed
Description
CHECK CALIBRATION
Measurement result is negative probably because of
shorter test leads than calibrated (negative value is higher
than 5 digits).
Calibrate test leads again!
EXTERNAL VOLTAGE
External voltage higher than 3 V is applied between two
R
PE
or between two SENSE test terminals
(measurement is not running) or higher than 10 V
(measurement is running).
External voltage higher than 5 30 V is applied between
any R
PE
or SENSE test terminal and GND.
Remove external voltage!
LIMIT OUT OF RANGE
Calculated limit value is < 1 (EN60204 SET Z limit mode)
FUSE F4!
Fuse F4 is blown.
ERROR 1!
Internal fuse may be blown!
The fuse is not customer replaceable, send the tester into
a service department.