HT instruments FULLTEST3 User Manual
Page 87

FULLTEST
3
EN - 86
TRMS Load Current (POWER)
Display range
A
Resolution
A
Accuracy
Overvoltage protection
0.00
0.99
0.01
± (3% rdg. + 3 dgt)
CAT II 300 V
1.0
20.0
0.1
Frequency range:
15
723 Hz
Over-range limit:
Yes, the measurement will be automatically interrupted 10 s
after exceeding 16 A
Pre-Test
Grounding of PE terminal on wall schuko socket
TRMS Mains Voltage (POWER)
Display range
V
Resolution
V
Accuracy
Overvoltage protection
195
253
1
± (2% rdg. + 2 dgt)
CAT II 300 V
Frequency range:
15
723 Hz
Apparent Power / Active Power (POWER)
Display range
VA/W
Resolution
VA/W
Accuracy
Overvoltage protection
0.0
99.9
0.1
± (5% rdg. + 10 dgt)
CAT II 300 V
100
999
1
± (5% rdg. + 3 dgt)
1.00
5.06 k
10
Limit value P
APP:
Adjustable
6
999 VA, 1.00 kVA 5.06 kVA
Power factor (POWER)
Display range
Resolution
Accuracy
Overvoltage protection
0.00
1.00
0.01
Respect the accuracy of P
APP
and P
ACT
CAT II 300 V
TRMS Leakage Current I
PE
(POWER)
Display range
Resolution
Accuracy
Overvoltage protection
0.25
19.99 mA
0.01 mA
± (3% rdg. + 3 dgt)
CAT II 300 V
20.0
49.9 mA
0.1 mA
0.05
0.99 A
0.01 A
1.0
10.0 A
0.1 A
Load current influence:
± 0.01 mA / A
Limit value:
Adjustable 0.25 mA
10.00 A, standard limit value 3.50 mA
Frequency range:
40 Hz
100 kHz (characteristics according to
DIN EN 61010-1 Annex A, fig. A.1 and VDE 0411 Part 1)
Measurement method:
Differential
Mains cord polarity exchange
Internally by using the L POS key
UUT mains on switch test:
Yes, the test done by measuring L current, limit value 25 mA
Over-range limit (I
PE
):
Yes, the measurement will be automatically interrupted 10 s
after exceeding 10 A
Over-range limit (I
L
):
Yes, the measurement will be automatically interrupted 10 s
after exceeding 18 A