HT instruments FULLTEST3 User Manual
Page 82

FULLTEST
3
EN - 81
Dielectric withstanding test (DIELECTRIC)
Nominal test
voltage U
N
(V)
Output
Resolution
(V)
Accuracy of
output voltage
Overvoltage
protection
250
800
COM & 0.25
0.80 kV
10 ±
3%
U
N
CAT III 300 V
810
2500
COM & 0.81
2.50 kV
2510
5100
COM & 2.51
5.10 kV
Nominal test voltage U
N
: Adjustable
250
5100 V, 50/60 Hz (floating) in steps of 10 V
Distortion of test voltage:
Crest factor = 1.414 ± 5%
Measurement modes:
MANUAL, RAMP (timer) or BURN
Output power:
500 VA at 5100 V
Leakage current I
APP
:
Display range
(mA)
Resolution
(mA)
Accuracy
0
200
1
± (3% rdg + 2 mA)
Leakage current I
REAL
:
Display range
(mA)
Resolution
(mA)
Accuracy
0
110
1
± (3% rdg + 4 mA)
Nominal breaking current (I
APP
or I
REAL
): Adjustable 1
110 mA in steps of 1 mA
Short-circuit current:
> 200 mA
Breaking time:
< 30 ms
Note!
Always connect COM terminal to GND if measured OUT is grounded, otherwise possible capacitive
leakage current may flow to ground which may disturb the measurement!