HT instruments FULLTEST3 User Manual

Page 85

background image

FULLTEST

3

EN - 84

Global earth resistance without tripping RCD (R

A

)

I

N

(mA)

Display range R

A

(

)

Resolution

(

)

Accuracy

Overvoltage

protection

10

0

 2000

1

± (3% rdg. + 1

), 15

 2000

*

CAT III 300 V

30

0.0

 99.9

0.1

± (3% rdg. + 1

), 10

 2000

*

100

 2000

1

100

0.0

 99.9

0.1

± (3% rdg. + 3 dgt)*

100

 1000

1

300

0.0

 99.9

0.1

100

 300

1

500

0.0

 99.9

0.1

100

 200

1

650

0.0

 99.9

0.1

100

 150

1

1000

0.0

 100.0

0.1

* The accuracy may be affected unstabile mains voltage!

Test current:

I

N

/2

Input voltage range:

100

265 V, 50/60 Hz

Nominal mains voltage :

230 or 240 V

Mains voltage U

L/PE

:

Display range

(V)

Resolution

(V)

Accuracy

100

 265

1 ±

(3%

rdg.)

Input resistance (U

L/PE

): 450

k


Contact voltage U

C

at I

N

:

Display range

(V)

Resolution

(V)

Accuracy

0

 100 (U

C LIM

= 50 V)

1

± (3% rdg. + 3 V)

0

 50 (U

C LIM

= 25 V)