beautypg.com

HT instruments FULLTEST3 User Manual

Page 89

background image

FULLTEST

3

EN - 88

TRMS Leakage current using current clamp (I

LEAK

)

Range

Display range

Resolution

Accuracy

(w/o clamp error)

Overvoltage protection

1000 mA

0.0

 99.9 mA

0.1 mA

± (3% rdg. + 3 dgt)

One measurement

terminal grounded

100

 1000 mA

1 mA

100.0 A

0.00

 9.99 A

0.01 A

10.0

 100.0 A

0.1 A

1000 A

0.0

 99.9 A

0.1 A

100

 1000 A

1 A

Input voltage range:

0

 1 V AC

Input resistance:

1 M

Frequency range:

40 Hz

 100 kHz (characteristics according to

DIN EN 61010-1 Annex A, fig. A.1 and VDE 0411 Part 1)

Clamp type:

AC, output voltage 1 V / measuring range, type

HT96U (measuring ranges 1 A, 100 A, 1000 A)

LIM value (1000 mA range):

Adjustable 0.1 ... 99.9 mA, 100 ... 1000 mA

LIM value (100.0 A range):

Adjustable 0.1 ... 100.0 A

LIM value (1000 A range):

Adjustable 1 ... 1000 A


TRMS Leakage Current on test socket (I

LEAK

)

Display range

Resolution

Accuracy

Overvoltage protection

0.25

 49.99 mA

0.01 mA

± (3% rdg. + 3 dgt)*

CAT II 300 V

0.05

 0.99 A

0.01 A

1.0

 10.0 A

0.1 A

* Clamp error not included

Load current influence

± 0.01 mA / A

Limit value

Adjustable 0.25

 19.99 mA, 20.0  49.9 mA, 0.05  0.99 A,

1.0

 10.0 A

Frequency range

40 Hz

 100 kHz (characteristics according to

DIN EN 61010-1 Annex A, fig. A.1 and VDE 0411 Part 1)

Measurement method

Differential

Mains cord polarity exchange

Internally by using the L POS key

UUT mains ON switch test

Yes, the test done by measuring L current, limit value 25 mA

Over-range limit (I

PE

)

Yes, the measurement will be automatically interrupted 10 s

after exceeding 10 A

Over-range limit (I

L

)

Yes, the measurement will be automatically interrupted 10 s

after exceeding 18 A

Pre-Tests

- Grounding of PE terminal on wall schuko socket
- Status of ON/OFF mains switch on UUT (measurement of

power consumption on schuko test socket, limit value 6 VA)