GOM-802 DC MILLI-OHM METER
USER MANUAL
⎯ ⎯
1
1. PRODUCT INTRODUCTION
1-1. Description
GOM-802 is a high precision programmable DC Milli-ohm meter suitable
for the low resistance measurements of switches, relays, connectors, PCB
tracks and variety of other devices. With the easy-to-use features, superior
performance, and automatic test interfaces described as below, GOM-802 is
obviously a reliable and handy instrument for the low resistance
measurements.
Easy-to-use features:
Utilizing the arrow key for setting the percentage of upper and lower limit based on
the normal value can make it easier to execute the HI-LO-GO
comparator
function.
Besides, through the Alarm Buzzer setting can indicate the PASS/FAIL status, and
all kinds of output status can be indicated through Handler Interface output.
The RELATIVE feature that enables GOM-802 to remove the stray resistance easily,
and the 20 sets of Normal/Hi/Lo setting memories can satisfy with different kind of
test condition. Also, the last setting will be recalled every time when the GOM-802
is turned on.
Superior Performance:
There are nine measurement ranges from 30mΩ to 3MΩ selected automatically or
manually with the constant current between 1μA and 1A, 0.05% high accuracy,
1μΩ resolution and four terminals Kelvin connection to make a reliable and
consistent test result.
The flexible choice for a high measurement accuracy at speed of 7 sampling/sec
(full scale at 30000) or a less measurement accuracy at speed of 30 sampling/sec
(full scale at 3000) can match with the speed requirement of test.