GOM-802 DC MILLI-OHM METER

USER MANUAL

⎯ ⎯

14

1) Under Compare function, press [NORMAL] button to turn on/off

the function of normal value setting.

2)

When the normal value setting is enabled, the maximum number
displayed on the normal value area will be flashing, adjust the value
by using

[▲]and [▼], and move the flashing number by using [ ] or

[ ], then press [ENTER] to confirm the correct value. The maximum

normal value is set at 33000 while the minimum normal value is set at

00000.

3) After setting, the HI, LO, GO judgment is according to the new

setting value.

4) If the setting is not saved, once jump out the setting function and re-

power on, the setting will be cleared.

High limit percentage setting:

1) Under the compare function, press [HIGH] to turn on/off the up

limit percentage setting function, and press [SHIFT] to clear the
setting.

2) After setting, the maximum number displayed on the up limit

percentage area will be flashing, adjust the value by using

[▲]and

[▼], and move the flashing number by using [ ] or [ ], then press

[ENTER] to confirm the correct value. The maximum up limit percentage

value is set at 999 while the minimum up limit percentage value is set at

000.

3) After setting, the HI, LO, GO judgment is according to the new

setting value.

4) If the setting is not saved, once leave the setting function and re-

power on, the setting will be cleared.

GOM-802 DC MILLI-OHM METER

USER MANUAL

⎯ ⎯

15

Low limit percentage setting:

1) Under the compare function, press [LOW] to turn on/off the down

limit percentage setting function, and press [SHIFT] to clear the
setting.

2) After setting, the maximum number displayed on the up/down limit

percentage area will be flashing, adjust the value by using

[▲]and

[▼], and move the flashing number by using [ ] or [ ], then press

[ENTER] to confirm the correct value. The maximum down limit

percentage value is set at 999 while the minimum down limit percentage

value is set at 000.

3) After setting, the HI, LO, GO judgment is according to the new

setting value.

4) If the setting is not saved, once leave the setting function and re-

power on, the setting will be cleared.

The save function for the compare setting:

1) Under Compare function, set the high and low limit percentage of

normal value according to the setting procedure of Normal value,
High limit percentage and low limit percentage.

2)

Press [SHIFT] and [RECALL]

in sequence

getting into save and

recall function of Compare setting.

3)

Switch to save function by pressing

[▲] or [▼], and the “SAVE”

message will be displayed on the panel, press [ENTER] to confirm the

setting, or press [SHIFT] to clear the setting.

4)

Among the message “S MXX” displayed on the panel, XX means the

number of the storage group from 00 to 19. Use [▲] or [▼], [ ] or

[ ] to adjust the save location, then press [ENTER] to replace the

previous save location, or press [SHIFT] to clear the setting.