GOM-802 DC MILLI-OHM METER

USER MANUAL

⎯ ⎯

20

3)

Under the TRIG mode, change the range or power on the instrument, the

“————” message will be displayed on the measurement displayed area,

now the instrument is in the TRIG standby status. The decimal is

represented the range. When proceeding the TRIG measurement, the

measurement-displayed area will display the present measuring value.

(4) Handler interface

The handler interface provides a TTL input (START), 6 TTL outputs,

including LOW, HIGH, FAIL, PASS, EOT and READY (

This function is

valid only under the resistance measurement mode and the compare mode is

enabled)

.

9-PIN D-SHELL (FEMALE)

START: Start trigger and proceed one time measurement, negative edge

trigger.

READY: Output “1” to GND, means the measurement is completed and

can proceed next trigger.

EOT: Output “1” to GND, means the AD convert procedure is

completed, can change the DUT.

LOW: Output “1” to GND, means the judgment of the compare

function is low.

HIGH:

Output “1” to GND, means the judgment of the compare

function is high.

EOT

READY

HIGH

LOW

GND
+5V
PASS
FAIL

START

GOM-802 DC MILLI-OHM METER

USER MANUAL

⎯ ⎯

21

FAIL:

Output “1” to GND, means the judgment of the compare

function is High or low.

PASS:

Output “1” to GND, means the judgment of the compare

function is GO.

The output current from all output collector and +5V

terminal can not exceed 60mA.

(5) Interface setting

Please refer to the programmable manual.

(6) SCAN interface

1) The scan interface provides 6 TTL outputs, including READY,

PASS, LOW, HIGH, FAIL, CLOCK, and STRB.

9-PIN D-SHELL (FEMALE)

RELAY: Control the relay signal, and series signal output.

PASS:

Indicate PASS signal, and series signal output.

LOW:

Indicate LOW signal, and series signal output.

HIGH:

Indicate HIGH signal, and series signal output.

CLOCK: When every group of output signal (including READY,

PASS, LOW, HIGH) is ready, a CLOCK signal will be sent

out. There are 100 groups of output signals totally.

STRB: After all 100 groups of output signals have been ready, a

STRB signal will be sent out.

PASS

RELAY

CLOCK

STRB

GND
+5V
LOW
HIGH