measurement-displayed area will display the present measuring value.
(4) Handler interface
The handler interface provides a TTL input (START), 6 TTL outputs,
including LOW, HIGH, FAIL, PASS, EOT and READY (
This function is
valid only under the resistance measurement mode and the compare mode is
enabled)
.
9-PIN D-SHELL (FEMALE)
START: Start trigger and proceed one time measurement, negative edge
trigger.
READY: Output “1” to GND, means the measurement is completed and
can proceed next trigger.
EOT: Output “1” to GND, means the AD convert procedure is
completed, can change the DUT.
LOW: Output “1” to GND, means the judgment of the compare
function is low.
HIGH:
Output “1” to GND, means the judgment of the compare
function is high.
EOT
READY
HIGH
LOW
GND
+5V
PASS
FAIL
START
GOM-802 DC MILLI-OHM METER
USER MANUAL
⎯ ⎯
21
FAIL:
Output “1” to GND, means the judgment of the compare
function is High or low.
PASS:
Output “1” to GND, means the judgment of the compare
function is GO.
The output current from all output collector and +5V
terminal can not exceed 60mA.
(5) Interface setting
Please refer to the programmable manual.
(6) SCAN interface
1) The scan interface provides 6 TTL outputs, including READY,
PASS, LOW, HIGH, FAIL, CLOCK, and STRB.
9-PIN D-SHELL (FEMALE)
RELAY: Control the relay signal, and series signal output.
PASS:
Indicate PASS signal, and series signal output.
LOW:
Indicate LOW signal, and series signal output.
HIGH:
Indicate HIGH signal, and series signal output.
CLOCK: When every group of output signal (including READY,
PASS, LOW, HIGH) is ready, a CLOCK signal will be sent
out. There are 100 groups of output signals totally.
STRB: After all 100 groups of output signals have been ready, a
STRB signal will be sent out.
PASS
RELAY
CLOCK
STRB
GND
+5V
LOW
HIGH