GOM-802 DC MILLI-OHM METER

USER MANUAL

⎯ ⎯

16

The recall function for the compare setting:

1) Under Compare function, set the high and low limit percentage of

normal value according to the setting procedure of Normal value,
High limit percentage and low limit percentage.

2) Press [SHIFT] and [RECALL]

in sequence

getting into save and

recall function of Compare setting.

3) Switch to recall function by pressing

[▲] or [▼], and the “CALL”

message will be displayed on the panel, press [ENTER] to confirm the

setting, or press [SHIFT] to clear the setting.

4)

Among the message “C MXX” displayed on the panel, XX means the

number of the recall group from 00 to 19. Use [▲] or [▼], [ ] or [ ]

to adjust the recall location, the panel will display the normal value and

the up/down limit percentage of normal value.

5) After recall setting, the HI, LO, GO judgment is according to the

new setting value.

The Buzzer function setting:
1) Under the Compare mode, press [SHIFT] and [ ]

in sequence

getting into buzzer setting. The panel will display the present
setting status, the message “NON” means no action, “bP” means the
buzzer is activated when the test is pass, and “bF” means the buzzer
is activated when the test is failed.

2) Use

[▲] or [▼] to select “NON”, “bP” or “bF”, then [ENTER] to confirm

the setting, or press [SHIFT] to clear the setting.

3)

The buzzer will be disabled once leave the Compare mode.

Display measurement value percentage:
1) Under the Compare mode, press [SHIFT] and [%/VALUE]

in

sequence

to select resistance value which is based on actual value or

normal value to calculate the percentage, and displays it on the
panel.

Normal value:

20.000

Actual value: 10.000
Percentage:

050.00%

GOM-802 DC MILLI-OHM METER

USER MANUAL

⎯ ⎯

17

2) Enable the measurement value percentage calculation, and the auto

range function will be annulled.

(3) REL FUNCTION

1) Use this function to do zero adjustment when the test lead or the

resistance of the DUT needs to be adjusted.

2) Pre-set a reference value, then every value must minus the reference

value before displaying on the panel for the compare judgment.

3) Under the compare mode, enable REL function, the auto-range

function will be annulled.

4) After calculation, if the value is negative, the minus symbol and the

number will be displayed alternately on the position of MSD (most

significant digit) under low converted speed mode (7 times/second),

while only the minus symbol will be displayed on the position of

MSD under high converted speed mode.

6-2. Temperature measurement

(1) Temperature measurement

1) Connect an optional temperature probe to the rear panel.

2) Press [SHIFT] and [TEMP]

in sequence

getting into main function

selection mode.

3) Press

[▲] or [▼] can switch over 3 kinds of functions of “ohm”, “℃”, and

“TC” on the panel.

4) Select “ ℃ ”and press [ENTER] getting into temperature

measurement mode. The measuring value will be displayed on the

normal value displayed area.