GW Instek GOM-802 User Manual User Manual
Page 14
GOM-802 DC MILLI-OHM METER
USER MANUAL
⎯ ⎯
22
2) SCAN Operation
{ Set to SCAN mode from the resistance measurement mode, first
enable the compare function and change the range manually, then
press [SHIFT] and [SCAN] in sequence.
| Set sweep count by adjusting
[▲] or [▼], [ ] or [ ], the
maximum count can be up to number 100. Press [ENTER] to confirm
the setting or press [SHIFT] to cancel the setting.
}
Set one count sweep delay time up to 30000 unit maximum or 30 unit
minimum (unit: 16.2ms) by adjusting [▲] or [▼], [ ] or [ ], then
press [ENTER] to confirm the setting or press [SHIFT] to cancel the
setting.
~
When the setting is completed, the message of “READY” will be
displayed on the measurement value displayed area. Now the setting of
range and up/down limit of the compare function can not be changed.
Proceed sweep function by press [MANUAL TRIG], and the result will
be output through SCAN interface.
Press [SHIFT] and [SCAN]
in sequence
to leave SCAN mode.
After sweep, the measurement results can be observed by
adjusting
[▲] or [▼].
Please refer to the procedure of SCAN interface output as follows:
GOM-802 DC MILLI-OHM METER
USER MANUAL
⎯ ⎯
23
Step 1. After the SCAN setting, the message of “READY” will be
displayed on the panel, SCAN interface starts output:
RELAY:
PASS:
LOW:
HIGH:
CLOCK:
∬
∬
STRB:
Step 2. Press [MANUAL TRIG] to start scan:
RELAY:
∬
∬
PASS:
∬
∬
LOW:
∬
∬
HIGH:
∬
∬
CLOCK:
∬
∬
STRB:
∬
∬
1 2 3 4
99 100
1 2
99 100