PST-3201/3202 PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ 12 ⎯

Example: Set current at 1.000A.

Press [I SET][1][.][0][0][0][ENTER]
When the load current through output terminal exceeds

the setting value, the instrument is operated in the C.C.
mode, if not exceeds the setting value, the instrument is
operated in the C.V. mode.

5-2.Over Voltage /Current Protection Setting

At first, select the wanted channel by pressing [SHIFT][CHx] now
the cursor is set to CHx (x=1, 2 or 3).
--Over Voltage Protection Setting:

Method 1: Set OVP voltage level by pressing [OVP SET], and

using number key to key in [voltage value], then press
[ENTER].

Method 2: Press [OVP SET] and using knob to input [voltage

value], the OVP voltage level setting will be changed
immediately, then press [ENTER] to terminate the
OVP voltage level setting. Obviously, using this
method, the OVP voltage level will be changed
immediately following the input value through knob.

Example: Set OVP voltage at 33.00V.

Press [OVP SET][3][3][.][0][0][ENTER]

--OVP Status Clear Up:

When the output voltage exceeds 33.00V of the setting voltage,
the output of the instrument will be off and get into OVP mode by
displaying “Over Voltage Protection…” on the panel. Now press
[SHIFT][OVP RESET] to clear OVP status, back to previous
status.

--Over Current Protection Setting:

Turn on/off the OCP of each channel individually by pressing
[OCP]. If OCP is on, when the output current equals or exceeds

PST-3201/3202 PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ 13 ⎯

the current value setting, the output of the instrument will be off
and get into over current protection mode by displaying “Over
current protection” on the display. Press [OCP] to clear OCP
status, and back to previous status.

5-3. Voltage/Current Step Setting:

Press [SHIFT][STEP] getting into the item selection picture, using
the knob to set cursor to the setting item which you want to modify,
input directly the wanted value and press [ENTER]. Store the setting
by using the knob to move the cursor to [SAVE], then press
[ENTER] again to complete the setting and storing. If want to cancel
the setting, move the cursor to [EXIT] and [ENTER] with knob to
terminate the setting without storing.

Example:Set the step voltage of Channel 1 at 1.00V and the step

current at 0.10A.

Press [SHIFT][STEP], elect CH1 Voltage and input
[1][.][0][0][ENTER], then select CH1 Current and input
[0][.][1][0][0][ENTER]. Finally, using the knob to move the cursor
to [SAVE] and press [

ENTER] to complete the setting and storing.

Note: The setting of the Step voltage and Step current of CH1 CH2

CH3 can be proceeded at the same display window.

5-4.Information Storing and Recalling setting:

--Information Storing Setting:

Press [SHIFT][STORE] getting into item selection picture, using