PST-3201/3202 PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ 22 ⎯

(1) Make sure the power cord is unplugged.

(2) Change the AC selects switch to the desired line voltage position.

(3)

A change in line voltage may also require a corresponding

change of fuse value. Install the correct fuse value as listed on

rear panel.

6-3. ADJUSTMENT AND CALIBRATION

--Preparation

1. 30 minutes warm up before calibration.

2. Ambient temperature:23±5°C, Humidity: Under RH80%.

--Output Calibration Steps:

[Step 1]

Press [SHIFT][‧] to appear Password input window, input Password

(vary with different models: PST-3202: 3202, PST-3201: 3201) by

using the number key, press [ENTER] key. When the value has been

input by mistake, but not yet pressed [ENTER], can use the knob to

clear the input value.

PST-3201/3202 PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ 23 ⎯

[Step 2]

When get into Calibrated item selection window, set the cursor to the

calibrated item by using the knob and press [ENTER]. After the

calibrated item is selected, proceeding voltage, current and over-

voltage calibration step by step.

[Step 3] Voltage Calibration Steps

After the channel is selected, set the cursor to [voltage] by using the

knob and press [ENTER] getting into voltage calibration procedure.

[Step 3.1]

After the proper measured setting has been done according to the step

displayed in the window, press again [ENTER] to get into voltage

value input window.

[Step 3.2]

Input measured voltage value directly by using the number key, and

press [ENTER] to complete the measured voltage (MIN) input (When

the value has been input by mistake, but not yet pressed [ENTER],

can use the knob to clear the value.)

Note: The DMM selected for the measurement must have the

resolution of three digits of the decimal point at least (1mV).

Also the input value is to take two digits of decimal point of

effective value (10mV) and run off the rest.

[Step 3.3]

After the proper adjustment of measured value has been done

according to the step displayed in the window, press [ENTER] to

complete Maximum voltage adjustment. The DMM selected for the