PST-3201/3202 PROGRAMMABLE POWER SUPPLY
USER MANUAL
⎯ 26 ⎯
Note: Now, the calibrated data of channel 1 has not yet been stored. If
the calibration is aimed at channel 1 only, can set the cursor to [Save]
by using the knob and press [ENTER] to finish the procedure with
storing.
If want to cancel the calibration, using the knob to set the cursor to
[EXIT] and press [ENTER] to terminate the procedure without storing.
The calibration procedure as mentioned above can be aimed at
channel 2 and channel 3 only.
[Step 6]
Repeat above calibration steps on channel 2 and channel 3.
[Step 7] Parallel Calibration steps
Set the cursor to [Parallel] by using the knob and press [ENTER]
getting into parallel current calibration procedure.
[Step 7.1]
After the proper measured setting has been done according to the steps
displayed by the window, press again [ENTER] to get into the operand
procedure of parallel current calibrated value.
[Step 7.2]
After the operand procedure of Parallel current calibration value has
been completed and the proper measured setting has been done
according to the steps displayed in the window, press [ENTER] to get
into the window for Parallel Current (MAX.) calibrated value input.
PST-3201/3202 PROGRAMMABLE POWER SUPPLY
USER MANUAL
⎯ 27 ⎯
[Step 7.3]
Input Parallel current measured value directly by using the number key,
and press [ENTER] to complete the Parallel current (MAX) measured
value input (When the value has been input by mistake, but not yet
pressed [ENTER], can use the knob to clear the value.)
Note: The DMM selected for the measurement must have the resolution
of four digits of the decimal point at least (0.1mA). Also the input value
is to take three digits of decimal point of effective value (1mA) and run
off the rest.
[Step 7.4]
Press again [ENTER] getting into the window for Parallel current
calibrated value input.
[Step 7.5]
Input Parallel current measured value directly by using the number key,
and press [ENTER] to complete the Parallel current (MIN) measured
value input (When the value has been input by mistake, but not yet
pressed [ENTER], can use the knob to clear the value.)
Note: The DMM selected for the measurement must have the resolution
of four digits of the decimal point at least (0.1mA). Also the input value
is to take three digits of decimal point of effective value (1mA) and run
off the rest.