PST-3201/3202 PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ 16 ⎯

the value of End must be larger than Start’s.

5-6. Auto-operation Mode

--Delay Time Setting:

Press [DELAY] getting into item selection picture, using the knob
to set cursor to the wanted setting item and input the time directly
by using number key, and press [ENTER]. Finally using the knob to
set the cursor to [End] and press [ENTER] to complete the setting.
The Delay Time setting has to be further stored in the specific place
of the memory address according to the procedure of 5-4
Information Storing setting. Please note, when the storing
procedure is proceeding, all the other settings of the instrument will
be also stored in the same place of memory address.
If want to cancel the setting, set the cursor to [Exit] by using the
knob and press [ENTER] to terminate the setting without storing.

--Auto-operation

Press [SHIFT][RECALL] getting into item selection picture, using
the knob to set cursor to recall range and press [ENTER] to appear
the picture of auto operation setting, set the cursor to the item to be

PST-3201/3202 PROGRAMMABLE POWER SUPPLY

USER MANUAL

⎯ 17 ⎯

modified by using the knob and using the number key to input [auto
operation setting value] and press [ENTER]. After modification,
using the knob to set the cursor to [Save], and press [ENTER] to
complete the setting and storing. If want to cancel the setting, just
set the cursor to [Exit] by using the knob and press [ENTER] to
terminate the setting without storing.

Note: When input “00” repeatedly can select Cycle setting.

5-7. Parallel Operation Mode

Press [SHIFT][PARA] getting into parallel operation mode. In this
mode, output voltage and current are mainly operated in channel 2.
The settable range of output voltage is same as channel 2 while the
settable range of output current is two times of channel 2.

Example:

(1) Channel 1: Voltage=10V, Current= 1A.
(2) Channel 2: Voltage=20V, Current= 2A.
(3) Press [SHIFT][PARA] getting into parallel mode.

(4) Output voltage=20V, output current=4A.

5-8.Track Operation Mode

Press

[SHIFT][TRACK] getting into Track Operation Mode.

In this

mode, output voltage and current are mainly operated in channel 2.
The settable range of output voltage is same as channel 2, but the
output current can be set with different value.
Example:

(1)

Channel 1: Voltage=10V, Current= 2A.