beautypg.com

6 firmware upgrade – Detcon 1600A-N4X User Manual

Page 30

background image

1600A-N4X

1600A-N4X Instruction Manual

Rev. 0.2

Page 26 of 38

Bit 7

1 = Channel 8 Cal

0 = Channel 8 no Cal

Bit 6

1 = Channel 7 Cal

0 = Channel 7 no Cal

Bit 5

1 = Channel 6 Cal

0 = Channel 6 no Cal

Bit 4

1 = Channel 5 Cal

0 = Channel 5 no Cal

Bit 3

1 = Channel 4 Cal

0 = Channel 4 no Cal

Bit 2

1 = Channel 3 Cal

0 = Channel 3 no Cal

Bit 1

1 = Channel 2 Cal

0 = Channel 2 no Cal

Bit 0

1 = Channel 1 Cal

0 = Channel 1 no Cal

5.6 Firmware Upgrade


The 1600A-N4X is capable of firmware upgrades for both the PLC and touch screen device. Upgrades will be
performed by the user. The following components are needed to perform firmware upgrades:

o

ProSave Software (Download free from

http://www.siemens.com

)

o

S7-200 USB PPI cable

o

S7-200 Memory cartridge

o

DB9 Gender Changer

Figure 33 S7-200 USB/PPI Cable

5.6.1 Touch Screen Upgrade

1. Disconnect blue DB9 cable from PLC Communications Module (Figure 34).
2. Using a Gender Changer, connect the USB/PPI cable to the DB9 Cable disconnected from the PLC

Communications Module, and connect the USB end of the USB/PPI cable to a PC or laptop.

3. From the touch screen display, switch to transfer mode by selecting:

Program Menu Ź Screen Utilities Ź Transfer

4. Open ProSave software on PC or laptop:

Start Ź All Programs Ź Simatic Ź ProSave Ź ProSave

5. Click on the “General” tab.
6. Under "Device Type", select TP 177A 6”.
7. Under "Connection", select Serial (via USB-PPI cable) for connection.
8. Click on the “Restore” tab.
9. Select the firmware file named “Detcon_1600AN1P_HMI_vXX.psb”.

XX represent current version.

10. Ensure that “Booting” check box is unchecked.
11. Press “Star Restore” to start the firmware upgrade.

NOTE: This operation may take some time, depending on the type of connection selected.

NOTE: Before entering Program Menu, an appropriate username and password may be required.