4.1.1 Reset Camera
Unlocking the Calibrate tab will also unlock the “Reset Camera” option in the Help menu. Reset
Camera can be used to reset the various features in the camera. You may want to reset the
camera if it is changed to an unusable state.
To reset the camera:
1. Click Help on the Menu Bar.
2. Select Reset Camera from the drop down menu.
3. Click Yes to reset the camera.
4. To complete the reset, turn the camera off. When the camera is turned on again, it will
reset its settings.
4.2 Adjusting Signal Settings
There are three settings and one tool in the Signals Section of the Calibrate Tab: Integration
Time, Gain, Baudrate, and Correct Image.
4.2.1 Integration Time
Adjusts the amount of signal the camera receives. Increasing the value can help increase video
contrast, and decreasing the value can help lower the signal if the image appears too bright.
To adjust the integration time:
1. Click Help and select Advanced from the drop down menu.
2. Enter the password: engineer
3. Click on the Calibrate Tab.