8 Troubleshooting

Before sending the pyrometer for repair, try to find the error and to solve the problem with the
help of the following list.

Temperature indication too low

Incorrect alignment of the pyrometer to the object

New correct alignment to achieve the max. temperature signal

Measuring object is smaller than spot size

Choose correct measuring distance

Measuring object is not always in the measuring spot of the pyrometer (e.g. swinging
wire or pouring stream)

Use max. value storage

Emissivity set too high

Set lower correct emissivity corresponding to the material

Lens contaminated

Clean lens carefully

Temperature indication too high

Emissivity set too low

Set higher correct emissivity corresponding to the material

The measurement is influenced by reflections of hotter machine parts

Try to avoid the influence of the interfering radiation or change measuring position

Measuring errors

Indicated temperature is decreasing during the use of the pyrometer, contamination of
the lens

Clean lens

Air contamination in the sighting path between pyrometer and object

Change position of the pyrometer with a clean sighting path

Strong HF-interferences

Change position of the pyrometer

Note: The wavelength band of the pyrometers reacts at low measuring temperatures
(below 130 °C) to incandescent lamps or very bright daylight (not valid for fluorescent
tube). For a correct measurement strong external light to the measured object should
be avoided.