Before sending the pyrometer for repair, try to find the error and to solve the problem with the

help of the following list.

Temperature indication too low

Incorrect alignment of the pyrometer to the object

New correct alignment to achieve the max. temperature signal

Measuring object is smaller than spot size. If needed, exchange optical head.

Choose correct measuring distance

Measuring object is not always in the measuring spot of the pyrometer (e.g. swinging
wire or pouring stream)

Use max. value storage

Emissivity set too high

Set lower correct emissivity corresponding to the material

Lens contaminated

Clean lens carefully

Temperature indication too high

Emissivity set too low

Set higher correct emissivity corresponding to the material

The temperature of fiber and optical head is not at least 30 °C lower than the measuring

temperature.

Use cooling jacket with air or water cooling

The measurement is influenced by reflections of hotter machine parts

Try to avoid the influence of the interfering radiation or change measuring position

Measuring errors

Indicated temperature is decreasing during the use of the pyrometer, contamination of

the lens

Clean lens

Air contamination in the sighting path between pyrometer and object

Change position of the pyrometer with a clean sighting path

Strong HF-interferences

Change position of the pyrometer

Instrument overheated

Use cooling jacket with air or water cooling

Temperature Indication is fluctuating, probably caused by changing emissivity

Wrong pyrometer type, use of ratio pyrometer recommended

Laser targeting light

Laser targeting light fails

Instruments max. temperature is exceeded. Use cooling jacket

Note: The wavelength band of the pyrometers reacts at low measuring temperatures
(below 130 °C) to incandescent lamps or very bright daylight (not valid for fluorescent

tube). For a correct measurement, strong external light to the measured object should
be avoided.