IS 50-LO plus / IGA 50-LO plus Manual

Instrument Settings 20

2

If the converter cover is opened the laser targeting light can be switched with
the button with the symbol. When the laser targeting light is switched on, the
display shows “PILT”.

:

With the arrow keys and all parameter settings can be displayed. They are

active after pushing the PAR button. Pushing the button longer changes the
settings in fast mode.

3 ESC / ENT: Pushing the ESC button changes the pyrometer to measuring mode. If a

parameter is changed with the arrow keys the indication of the ESC button
changes to ENT. Pressing the button again confirms the value into the

pyrometer. Changing the parameters again by pushing the PAR button doesn’t
confirm this value in the pyrometer. If no button is pressed for 30 s the
pyrometer changes to the temperature indication without accepting the
changed value.

4.2 Factory settings

Emissivity (

Emi) = 100%

Exposure time (t

90

) = min

Clear time (t

Clear

) = off

Analog output (mA) = 0 … 20 mA
Sub range (from / to) same as temperature range

Address (Adr) = 00
Baud rate (Baud) = 19200 Bd
Temperature display (C / F) = °C
Wait time (t

w

) for RS485 = 10

Interface (RS485 / RS232) = RS232