Calibration

13

Complete the following steps (refer to Figure 2-2 on page 4 and Figure 3-1 on page 8):

1. Open the

VENT

valve (8) and close the

COARSE ADJUSTMENT

valve (2).

2. Using the

RANGE SELECT

switch (19), select the lowest pressure range on the UPC5000/UPC5010.

3. Select the

ZERO/SPAN

position on the Condec Calibration Module (PN 60109). The

UPC5000/UPC5010 displays three dashes on the left side of the display to indicate this mode.

4. Depress the

ENTER

button on the module. The UPC5000/UPC5010 display reads zero. Repeat this step

for the mid and high ranges.

5. Close the

VENT

valve (8). Select the lowest pressure range of the UPC5000/UPC5010.

6. Turn the

VERNIER

(13) of the UPC5000/UPC5010 counter-clockwise to create a vacuum until the

display reads -0.05 negative offset.

7. Depress the

ENTER

button on the module. The UPC5000/UPC5010 display reads zero. Repeat this step

for the mid and high ranges.

8. Select

DATA RECALL

on the module. The display reads

1XXX

(three digits).

9. Press the

STORE

button on the module. The display reads

1 020

or

1 377

when the button is pressed.

3.9

Normal Mode Test

After completing the above calibration procedures, you must perform a normal mode test. A current generator
capable of generating 20 mA must be connected to the COMMON and CURRENT INPUT jacks, see Figure 2-2
(14). The DISPLAY SELECT switch (16) should be on the CURRENT position.

1. Set the Condec Calibration Module to the

NORMAL MODE

position.

2.

DISPLAY SELECT

switch should still be in the

CURRENT

position. Display will read

20.000

.

3. Turn the

DISPLAY SELECT

switch to the

VOLTAGE

position. Display will read

100.00

.

4. The pneumatic portion of the calibration is now complete and the pressure standard and the module can

now be disconnected.

3.10 Self-Check Test

Complete the following steps (see Figure 2-2).

1. Remove the CONDEC Calibration Module.

2. Depress the ZERO push-button (12) on the UPC5200/UPC5210 until the unit shows "

CAL

". The display

will show "

100

" to verify the unit's accuracy, then it will return to the normal mode automatically.

3. The pneumatic portion of the calibration is now complete and the pressure standard and the CONDEC

Calibration Module can now be disconnected.