UPC5000/UPC5010 Operation & Maintenance Manual
Figure 3-3. Linearity and Hysteresis Calibration
Starting with the instrument’s lowest pressure range, sequentially perform the thirteen steps described in
Table 3-2, for each pressure range being calibrated. Perform the following for each step:
1. Adjust input pressure to the appropriate value without overshooting the setting. If value is overshot, vent
unit and repeat steps.
2. Perform the action as indicated when the readings are stable. On units below 2000 PSI it should not take
longer than five minutes. Units above 2000 PSI and all absolute only units should take no longer than
fifteen minutes. If it is taking longer, check system for leaks. If no leaks are found, the CPU or transducer
may be defective.
When Step 11 is reached, the display changes so that the left most status symbol is
H
. This remains for Step 12
and down to approximately 0.00 PSI.
Step
Input Pressure % of Range
Operator Action Required
Status Symbol in
Left-most Digit
Remarks
1
0 (see note 4 below)
Press ZERO switch
L
Zero on display
2
10
Press ENTER button
L
Notes 1 & 2 below
3
20
Press ENTER button
L
Notes 1 & 2 below
4
30
Press ENTER button
L
Notes 1 & 2 below
5
40
Press ENTER button
L
Notes 1 & 2 below
6
50
Press ENTER button
L
Notes 1 & 2 below
7
60
Press ENTER button
L
Notes 1 & 2 below
8
70
Press ENTER button
L
Notes 1 & 2 below
9
80
Press ENTER button
L
Notes 1 & 2 below
10
90
Press ENTER button
L
Notes 1 & 2 below
11
100
No Action Required
H
Note 3 below
12
50
Press ENTER button
H
Notes 1 & 2 below
13
0 (see note 4 below)
No Action Required
L
Table 3-2. Linearization and Hysteresis Calibration Sequence