Page 14

background image

Calibration

11

NOTES:

1.

If reading is in motion or correction required is not within ±0.8% of full-scale, no entry is made.

2.

If entry is valid, the display momentarily indicates the correction value (in percent) and the memory location at
which it is stored.

3.

If 100% ±0.05% is not obtained, repeat the zero/span calibration sequence.

4.

Absolute only unit: Maximum vacuum standard display reading of 0.04 PSIA.

3.5

Shunt Resistor Calibration

To place the UPC5000/5010 into shunt calibration mode, install the Condec Calibration Module (PN 60109) and
select the

SHUNT MODE

position of the rotary switch. The display is shown in Figure 3-4.

Figure 3-4. Display in Shunt Resistor Calibration Mode.

With the UPC5000/UPC5010’s highest pressure range selected, perform the four step sequence described below.

1. Gage Only and Absolute/Gage Units: be sure the input pressure is set at 0 PSIG.

NOTE:

Absolute Only Units: Must use a vacuum pump with PSIA display, to reach as close to 0 PSIA as possible

(maximum vacuum standard display reading of 0.04 PSIA).

2. Press and hold the

ZERO

button on the module until a stable zero indication is obtained.

3. Release the

ZERO

button and allow the display to stabilize at its shunt resistor calibration number (100 ±

5.00%).

4. Press the

ENTER

button on the module. When accepted, the bottom half of all display digits

momentarily illuminate.

3.6

Voltage/Current Input Calibration

To calibrate a current generator capable of generating 20 mA, it should be connected to the

COMMON

and

CURRENT

INPUT

jacks (Figure 2-2 on page 4 [14]). The

DISPLAY SELECT

switch (16) should be in the

VOLTAGE

position.

1. Set the Condec Calibration Module (PN 60109) to the

ZERO/SPAN

position (see Figure 3-2 on page 9 for

display reading).

2. Press the

ENTER

button on the module. The display reads

0.00

.

3. Set the current generator for 20 mA output. Press the

ENTER

button on the module. The display should

read

100.000

.

4. Turn the

DISPLAY SELECT

switch (16) to the

CURRENT

position. Display will read 20.000.

5. Disconnect the current generator.

NOTE:

If the display reading is off, set the Current Generator to 0, and press the ENTER button on the Condec

Calibration Module. Set the Current Generator for 20 mA output. The display will read 20.000. If the display reading is
off, press the ENTER button on the module. If the display reading is not 20.000, CPU is faulty and requires servicing.