beautypg.com

3 zero/span calibration, 4 linearity and hysteresis calibration – Condec UPC5010 User Manual

Page 12

background image

Calibration

9

3.3

Zero/Span Calibration

Selecting the

ZERO/SPAN

position on the Condec Calibration Module (PN 60109) places the instrument into its

ZERO/SPAN

calibration mode. The display is shown in Figure 3-2.

NOTE:

Absolute Only Unit requires vacuum pump with PSIA indicator to obtain readings below local barometric

pressure.

Figure 3-2. Zero/Span calibration for Absolute/GageUnit

Starting with the instrument’s lowest pressure range, perform Steps 1 and 2 shown in Table 3-1 for each pressure
range.

Note:

Perform Step 1 in all ranges prior to doing Step 2.

Perform the following for each step:

1. Gage Only or Absolute/Gage Unit: Adjust input pressure to the appropriate (either 0 or 100%) value.

Absolute Only Units: Must use a vacuum pump with PSIA display, to reach as close to 0 PSIA as
possible.

2. Perform the action indicated in Table 3-1 when pressure input readings are stable.

NOTES:

1.

If readings are not stable or are not within ±20% of zero, the zero correction can’t be entered.

2.

If readings are not stable or are not within ±5% of 100%, the span correction cannot be entered.

3.

Absolute only unit: Maximum vacuum standard display reading of 0.04 PSIA.

3.4

Linearity and Hysteresis Calibration

Install the Condec Calibration Module (PN 60109) and select the

LYN/HYS

position of the rotary switch on the

module. This places the UPC5000/UPC5010 into its linearization/hysteresis calibration mode. The display is
shown in Figure 3-3 below.

NOTE:

The zero/span calibration needs to be performed prior to linearity and hysteresis calibration. For Absolute Only

Unit, vacuum pump with PSIA indicator must be used to obtain readings below local barometric pressure.

Step No.

Pressure Input Value

Operator Action Required

Resulting Display Indication

Remarks

1

0% (see Note 3
below)

Press ENTER

button

0%

Note 1 below

2

100%

Press ENTER button

100%

Note 2 below

Table 3-1. Zero and Span Calibration Sequence

This manual is related to the following products: