beautypg.com

Timer, event action > timer, Timer, event action > match time timer, Timer, event action > evnet action – Yokogawa Button Operated MV2000 User Manual

Page 45

background image

Timer, Event action > Timer

Setting

Selectable Range or Choices

Default Value

Timer No.

1/2/3/4

1

Timer

Mode

Off/Relative/Absolute

Off

Mode = Relative
Interval

Numeric value (00:01 to 24:00)

01:00

Reset at Math Start

On/Off

On

Mode = Absolute
Interval

1min/2min/3min/4min/5min/6min/10min/12min/

15min/20min/30min/1h/2h/3h/4h/6h/8h/12h/24h

1h

Ref.time

Numeric value (0 to 23)

0

Timer, Event action > Match time timer

Setting

Selectable Range or Choices

Default Value

Timer number

1/2/3/4

1

Match time timer

Kind

Off, Day, Week, Month, Year

Off

Kind > Day

Day

Numeric value (1 to 28)

1

Hour:Minute

Numeric value (00:00 to 23:59)

00:00

Timer action

Single/Repeat

Repeat

Kind > Week

Day of the

week

SUN/MON/TUE/WED/THU/FRY/SAT

SUN

Hour:Minute

Numeric value (00:00 to 23:59)

00:00

Timer action

Single/Repeat

Repeat

Kind > Month

Day

Numeric value (1 to 28)

1

Hour:Minute

Numeric value (00:00 to 23:59)

00:00

Timer action

Single/Repeat

Repeat

Kind > Year

Month

1/2/3/4/5/6/7/8/9/10/11/12

1

Day

Numeric value (1 to 31)

1

Hour:Minute

Numeric value (00:00 to 23:59)

00:00

Timer action

Single/Repeat

Repeat

Timer, Event action > Evnet action

Setting

Selectable Range or Choices

Default Value

Logic box number

1/2/3/4/.../39/40

1

Event-Action

Event

None/Remote/Relay/Switch/Timer/

MatchTimeTimer/Alarm/UserKey

None

Event = Remote

Remote number

1/2/3/4/5/6/7/8

1

Action

MemoryStart/Stop, MemoryStart,

MemoryStop, Trigger, AlarmACK, MathStart/

Stop, MathStart, MathStop, MathReset,

SaveDisplay, SaveEvent, Message, Snapshot,

DisplayRate1/2, ManualSample, TimerReset,

DisplayGroupChange, Flag, PanelLoad,

TimeAdjust, DisplayFreeze/Active, MemoryClear

DisplayGroupChange

Event = Relay

Relay number

I01/…/I06, I11/…/I16

I01

Action

MemoryStart/Stop, MemoryStart, MemoryStop,

Trigger, MathStart/Stop, MathStart, MathStop,

MathReset, SaveDisplay, SaveEvent, Message,

Snapshot, DisplayRate1/2, ManualSample,

TimerReset, DisplayGroupChange, Flag

DisplayGroupChange

Default MV2000 Settings

45

IM MV2000-02E

This manual is related to the following products: