Description of the Multi Batch Function
1
2
3
Index
Recording Display Data and Event Data
The following recording conditions are common to all batch groups. For details on the
recording operation, see section 1.4 in the User’s Manual.
Item
Description
Scan interval
1 second, 2 seconds, or 5 seconds. The same scan interval is used for
all batch groups.
Data record
One of either display data or event data. The same type of data is
recorded for all batch groups.
Display data
The same display data sampling interval (trend interval) is used for all
batch groups.
Event data
Use Free mode when recording event data. You cannot select Trigger
mode.
The same event data sampling interval (sample rate) is used for all
batch groups.
Recording Manual Sampled Data and Report Data (/M1 and /PM1 options)
The settings and behavior of these features are common to all batch groups. For details
on manual sampled data and report data, see section 1.4 in the User’s Manual.
Saving Measured Data to Storage Media
This section briefly explains how the DX saves various types of measured data to storage
media. For shaded entries, the DX performs the same operation as a DX without the
Multi Batch function. For details on saving measured data to storage media, see section
1.4 in the User’s Manual.
Item
Description
Auto save
Display data
Saves the data from all batch groups to the specified save
destination directory. The same file save interval for display
data is used for all batch groups.
Event data
Saves the data from all batch groups to the specified save
destination directory. The same file save interval for event
data (data length) is used for all batch groups.
Manual sampled
data
Saves the data to the specified save destination directory.
Report data
Saving data by
pressing FUNC
Save display
data
Saves the source batch group’s display data that the DX is
currently recording to files on the inserted CF card.
Save event data Saves the source batch group’s event data that the DX is
currently recording to files on the inserted CF card.
Manual save
Display data
Saves the data from all unsaved batch groups to the
specified save destination directory.
Event data
Manual sampled
data
Saves the unsaved data to the specified save destination
directory.
Report data
Saving data using keys
This operation is applicable when you insert a USB flash
memory device or when you display the DX’s memory
summary and save data. For details, see section 3.10 in
this manual or section 4.8 in the User’s Manual.
Snapshot
Saves the screen image data to the inserted CF card.
Saving data via
FTP
Display data
and event data
Transfers the data for all batch groups to the specified initial
path.
Report data
Transfers the data to the specified initial path.
Snapshot data
Media FIFO
The DX performs a FIFO operation only when the following
files are saved.
Display or event data files for all batch groups, manual
sampled data files, report files, and snapshot data files.