Yokogawa DO402 Dissolved Oxygen Analyzer User Manual
Page 92

IM 12J05D02-01E
11-4 Appendix 2
4/4
QIS 12J05D02-01E
3.7 Atmospheric Pressure Indication Check
Following Section 3.6, press the [ENT] key until the message display shows “PRESS.”
Check that the data display indication at this time is within the range from 900 (hPa) to
1100 (hPa).
Press the [ENT] key. The message display will show “READY.”
Continue to press the [ENT] key to return to normal measurement mode.
CONTACT
FAIL S3 S2 S1
3 2 1
32
71 72 51 52 41 42 31
11
T
1
12
T
2
61
62
65 66
15
RE
13
IE
17
+
-
+
-
mA 1
mA 2
Figure 1 Testing Circuit and Test Equipment
DO402G Converter
Decade
Resistance Box
Power Supply
Standard Current
Generator
+
-
SW1
+
-
DC Ammeter
Current Output 1
300
1%
+
-
DC Ammeter
300
1%
Current Output 2
See also other documents in the category Yokogawa Sensors:
- EJA120A (47 pages)
- EJA120A (79 pages)
- EJA130A (2 pages)
- EJA130A (4 pages)
- EJA120A (31 pages)
- EJA130A (47 pages)
- EJA120A (40 pages)
- EJA438 (5 pages)
- EJA120A (6 pages)
- EJA115 (85 pages)
- EJA210A (70 pages)
- EJA130A (4 pages)
- EJA430A (78 pages)
- EJA210E (52 pages)
- EJA210E (89 pages)
- EJA210E (170 pages)
- EJX120A (4 pages)
- EJA210E (9 pages)
- EJX115A (55 pages)
- EJA210E (41 pages)
- EJA210E (96 pages)
- EJX910A (9 pages)
- EJX910A (103 pages)
- FlowNavigator Software (163 pages)
- EJX910A (55 pages)
- EJX910A (175 pages)
- EJX910A (83 pages)
- EJA530A (67 pages)
- EJA120A (83 pages)
- EJX530A (52 pages)
- EJA110E (4 pages)
- EJA110E (85 pages)
- EJX120A (85 pages)
- EJA118 (76 pages)
- EJX118A (64 pages)
- EJA438 (72 pages)
- EJA430E (85 pages)
- EJX430A (76 pages)
- EJA430E (7 pages)
- EJX430A (6 pages)
- EJX430A (40 pages)
- EJA430E (41 pages)
- EJA430E (96 pages)
- EJX438A (10 pages)
- ADMAG AXR (194 pages)